First published · Chinese version: HY-IND232 仪表标定教程

You are here: Home > HY-IND232 calibration

HY-IND232 Weight Indicator Calibration: Step-by-Step Tutorial

This is the calibration procedure for the HY-IND232 series weight indicator, arranged from the company training video. Each step carries a frame from that moment in the clip and the time it appears at, so you can check it against the original recording.

Scope: this page is an operating procedure. A calibration result is not a legal conclusion; where local regulations apply, follow them.

If you get stuck during the procedure, or you want to confirm which version is on your site:

+86 183 5123 8137

One tap dials. Tell us the indicator model and what the site looks like.

What do I need before calibration?

These are the items that appear in the video captions and that the procedure actually uses:

The values in the video (0.01, 3 kg, 0.5 kg and so on) are worked examples: the captions themselves say we change it according to our own needs. They are not recommended values and not standard values.

What do the three training videos show?

The three clips below are the original company training videos, from the same batch of material as the frames on this page. The three versions differ as set out in the table further down; watch the clip that matches your indicator first, then work through the 30 steps.

Opening this page does not download a whole clip: only the information needed to play is loaded, and buffering starts when you press play. On a weak workshop signal, use the download link under each clip and keep that clip on the phone offline.

Black version HY-IND232 (physical keys Fn / T / 0 / ENTER) | 1280×720 | 1 min 44 s. The clearest of the three; the 30-step sequence on this page follows the order of its captions.

Download (7.5 MB)

Blue version HY-IND232+ (physical keys Fn / T / 0 / ENTER) | 720×404 | 2 min 0 s. The key path is close to the black version, but the level numbers are written FN1.1 / FN1.2 / FN1.3, and the full-scale example differs (the screen shows C03000.0). A few captions in this clip can only be read in part; each is marked where it occurs below.

Download (5.6 MB)

Silver version HY-IND232 (soft keys Fn / SP / CAL0, with the zero key) | 544×944 portrait | 2 min 5 s. The largest difference of the three: it takes the soft-key path, and an empty-scale check (noLAd) comes before loading test weights (AddLd1).

Download (7.2 MB)

The values and the engineering password demonstrated in the clips are worked examples: division, full scale and test weight all depend on the site; the engineering password follows the manual supplied with the unit and the factory setting, so do not copy it.

What are the 30 steps in the black-version video?

The steps below follow the 30 captions of the black-version calibration video in order; the number is the position of that caption in the clip. Each step gives three things: the caption text (only minimally reworded, meaning unchanged), its start and end time in the video, and a frame from the same interval.

The keys on this version are physical keys: Fn / T / 0 / ENTER (the captions write fn, T key, 0 key and confirm key respectively).

  1. IND232 indicator calibration (video title card, not an operating step)

    Black-version calibration video opening frame: the indicator shows 0.00 kg and the caption reads IND232 indicator calibration
  2. Press 0 and fn at the same time

    The indicator held in the hand ready for key presses; the caption reads press 0 and fn at the same time
  3. The indicator shows F1

    Indicator screen shows F1; the caption reads the indicator shows F1
  4. Press the confirm key

    A finger presses the ENTER confirm key; the caption reads press the confirm key
  5. Enter the engineering password (the screen shows P-00000)

    Indicator screen shows the P-00000 password entry screen; the caption is the password prompt
  6. T key shifts, 0 key increases the value

    Indicator screen shows part of the password entered; the caption reads the T key shifts and the 0 key increases the value
  7. Press the confirm key

    A finger holds an indicator key down; the caption reads press the confirm key
  8. The indicator shows F1.1

    Indicator screen shows F1.1; the caption reads the indicator shows F1.1
  9. Press the confirm key

    Indicator screen shows F1.1; the caption reads press the confirm key
  10. d 0.01 division selection

    Indicator screen shows the d 0.01 division screen; the caption reads d 0.01 division selection
  11. 0 key selects

    A finger presses the 0 key; the caption reads 0 key selects
  12. We take 0.01 as an example

    Indicator screen shows d 0.01; the caption reads we take 0.01 as an example
  13. Press the confirm key

    A finger presses the ENTER confirm key; the caption reads press the confirm key
  14. C 0003.00 vessel full scale

    Indicator screen shows the C0003.00 full-scale screen; the caption reads C 0003.00 vessel full scale
  15. We change it according to our own needs

    Indicator screen shows C0003.00; the caption reads we change it according to our own needs
  16. T key shifts, 0 key increases the value

    Indicator screen shows the full-scale value being changed; the caption reads the T key shifts and the 0 key increases the value
  17. We take 3 kg as an example

    Indicator screen shows C0003.00; the caption reads we take 3 kg as an example
  18. Press the confirm key

    A finger presses the ENTER confirm key; the caption reads press the confirm key
  19. F1.2 zero calibration

    Indicator screen shows F1.2; the caption reads F1.2 zero calibration
  20. Press the confirm key

    Indicator screen shows F1.2; the caption reads press the confirm key
  21. F1.3 load test weights

    Indicator screen shows F1.3; the caption reads F1.3 load test weights
  22. Press the confirm key after the test weights are loaded

    A finger presses the ENTER confirm key; the caption reads press the confirm key after the test weights are loaded
  23. Enter the test weight

    Indicator screen shows the test-weight entry screen; the caption reads enter the test weight
  24. T key shifts, 0 key increases the value (the caption here reads 0 key increases the number; it means the same as the earlier lines)

    Indicator screen shows the test-weight value; the caption reads the T key shifts and the 0 key increases the number
  25. We take 0.5 kg as an example

    Indicator screen shows 0.50; the caption reads we take 0.5 kg as an example
  26. Press the confirm key after the change

    A finger presses the ENTER confirm key; the caption reads press the confirm key after the change
  27. The indicator shows 0.5 kg

    Indicator screen shows 0.50 kg; the caption reads the indicator shows 0.5 kg
  28. Remove the test weights

    Indicator screen still shows 0.50; the caption reads remove the test weights
  29. The indicator returns to zero automatically

    Indicator screen back at zero, 0.00; the caption reads the indicator returns to zero automatically
  30. Calibration complete (the caption as shown; see the Scope note at the top of this page)

    Indicator screen shows 0.00; the caption reads calibration complete

How do the three versions differ?

The most important point: the key path is different. The silver version takes the soft keys Fn / SP / CAL0 (with the zero key); the black and blue versions take physical keys Fn / T / 0 / ENTER. The level numbers are written differently as well, so read the part that matches your model.

Black version HY-IND232 (physical keys Fn / T / 0 / ENTER)

Blue version HY-IND232+ (physical keys Fn / T / 0 / ENTER)

Blue-version indicator screen shows Fn1.1; the caption reads FN1.1 select the division

Blue version: FN1.1 selects the division (video at 00:33).

Blue-version indicator screen shows C03000.0; the caption reads C03000.0 vessel full scale

Blue version: the full-scale screen C03000.0 (video at 00:52). The caption here can only be read as vessel..., with the last two characters marked unreadable in the transcript; the black version writes the same sentence as vessel full scale, and both mean the same, so this page uses vessel full scale throughout.

Silver version HY-IND232 (soft keys Fn / SP / CAL0, with the zero key)

Silver-version indicator screen shows CAL0; the caption reads CAL0 selected with the FN key

Silver version: CAL0 selected with the FN key (video at 00:13). The soft-key path differs from the other two versions and is the largest difference of the three.

Key path and level numbers for the three versions
ItemBlack HY-IND232Blue HY-IND232+Silver HY-IND232
Enter the programPress 0 and fn togetherPress FN and the zero key togetherPress the zero key and fn together
Key typePhysical keysPhysical keysSoft keys
Confirm keyENTERConfirm keySP
Shift / increaseT shifts, 0 increasesT shifts, 0 increasesFN shifts, zero key increases
Level numbersF1.1 / F1.2 / F1.3FN1.1 / FN1.2 / FN1.3d → C → noLAd → AddLd1
Level for loading test weightsF1.3 load test weightsFN1.3 load test weights hereAddLd1 load test weights
Closing captionCalibration completeReturns to 0, calibration completeCalibration complete

What are the two key parameters?

These two sentences are written directly on the silver-version video captions, reproduced here:

Both sentences are written on the video frames and describe how the value is chosen for this indicator; the actual value still has to be set from the module configuration and the conditions on site.

What should I check during the procedure?

To check a step against the original clip, or if you are stuck on site:

+86 183 5123 8137

Still unsure which version you have?

Tell us the model on the nameplate, the keys on the front, and what the screen shows when you start.

Call +86 183 5123 8137