What do I need before calibration?
These are the items that appear in the video captions and that the procedure actually uses:
- A weight indicator that can enter the calibration program. The video shows the IND232 series (the black-version screen shows
IND232andWeight Indicator). - The engineering password. Entering the program shows
P-00000, and the engineering password00111is entered. The engineering password follows the manual supplied with the unit and the factory setting; it can differ between batches and years, so do not copy the value used here. - The division value (
d). Selected after entering the program; the video uses0.01for the black version and0.5for the blue version. - The vessel full scale (
C). The video showsC0003.00; the silver-version caption reads: the full scale is set from the maximum load of the equipment or vessel on site. - Test weights. Test weights are loaded at
F1.3, and after loading, their weight is entered into the indicator; the silver-version caption reads: the test weight is 10% to 20% of the total of the modules.
The values in the video (0.01, 3 kg, 0.5 kg and so on) are worked examples: the captions themselves say we change it according to our own needs. They are not recommended values and not standard values.
What do the three training videos show?
The three clips below are the original company training videos, from the same batch of material as the frames on this page. The three versions differ as set out in the table further down; watch the clip that matches your indicator first, then work through the 30 steps.
Opening this page does not download a whole clip: only the information needed to play is loaded, and buffering starts when you press play. On a weak workshop signal, use the download link under each clip and keep that clip on the phone offline.
Black version HY-IND232 (physical keys Fn / T / 0 / ENTER) | 1280×720 | 1 min 44 s. The clearest of the three; the 30-step sequence on this page follows the order of its captions.
Blue version HY-IND232+ (physical keys Fn / T / 0 / ENTER) | 720×404 | 2 min 0 s. The key path is close to the black version, but the level numbers are written FN1.1 / FN1.2 / FN1.3, and the full-scale example differs (the screen shows C03000.0). A few captions in this clip can only be read in part; each is marked where it occurs below.
Silver version HY-IND232 (soft keys Fn / SP / CAL0, with the zero key) | 544×944 portrait | 2 min 5 s. The largest difference of the three: it takes the soft-key path, and an empty-scale check (noLAd) comes before loading test weights (AddLd1).
The values and the engineering password demonstrated in the clips are worked examples: division, full scale and test weight all depend on the site; the engineering password follows the manual supplied with the unit and the factory setting, so do not copy it.
What are the 30 steps in the black-version video?
The steps below follow the 30 captions of the black-version calibration video in order; the number is the position of that caption in the clip. Each step gives three things: the caption text (only minimally reworded, meaning unchanged), its start and end time in the video, and a frame from the same interval.
The keys on this version are physical keys: Fn / T / 0 / ENTER (the captions write fn, T key, 0 key and confirm key respectively).
-
IND232 indicator calibration (video title card, not an operating step)
-
Press 0 and fn at the same time
-
The indicator shows F1
-
Press the confirm key
-
Enter the engineering password (the screen shows P-00000)
-
T key shifts, 0 key increases the value
-
Press the confirm key
-
The indicator shows F1.1
-
Press the confirm key
-
d 0.01 division selection
-
0 key selects
-
We take 0.01 as an example
-
Press the confirm key
-
C 0003.00 vessel full scale
-
We change it according to our own needs
-
T key shifts, 0 key increases the value
-
We take 3 kg as an example
-
Press the confirm key
-
F1.2 zero calibration
-
Press the confirm key
-
F1.3 load test weights
-
Press the confirm key after the test weights are loaded
-
Enter the test weight
-
T key shifts, 0 key increases the value (the caption here reads 0 key increases the number; it means the same as the earlier lines)
-
We take 0.5 kg as an example
-
Press the confirm key after the change
-
The indicator shows 0.5 kg
-
Remove the test weights
-
The indicator returns to zero automatically
-
Calibration complete (the caption as shown; see the Scope note at the top of this page)
How do the three versions differ?
The most important point: the key path is different. The silver version takes the soft keys Fn / SP / CAL0 (with the zero key); the black and blue versions take physical keys Fn / T / 0 / ENTER. The level numbers are written differently as well, so read the part that matches your model.
Black version HY-IND232 (physical keys Fn / T / 0 / ENTER)
- Enter the program: press 0 and fn at the same time → the indicator shows F1 → press ENTER
- Enter the password: screen
P-00000→ T key shifts, 0 key increases the value → press ENTER - Level numbers:
F1.1(division selection) →F1.2(zero calibration) →F1.3(load test weights) - Confirm key:
ENTER(the captions always write confirm key) - Example values in the video:
d 0.01→C0003.00(vessel full scale) → 3 kg → 0.5 kg - End: remove the test weights → the indicator returns to zero (caption: calibration complete)
Blue version HY-IND232+ (physical keys Fn / T / 0 / ENTER)
- Enter the program: press FN and the zero key at the same time → the display shows FN1 press (the caption here can only be read as shows FN1 press; the second half is marked unreadable in the transcript)
- Enter the password: screen
P-00000→ after the password is entered, press the confirm key to enter the program - Level numbers:
FN1.1(select the division) →FN1.2(check the vessel, marked unreadable in the transcript) →FN1.3(load test weights here) - Confirm key: the confirm key
- Example values in the video:
d 0.5→C03000.0(caption: vessel full scale) → 2000.0 → 1000.0 - End: returns to 0 automatically (caption: returns to 0 automatically, calibration complete)
- This clip has the lowest resolution (720×404), so a few captions can only be read in part, and the transcript marks each one as unreadable. This page writes only what was read.

Blue version: FN1.1 selects the division (video at 00:33).

Blue version: the full-scale screen C03000.0 (video at 00:52). The caption here can only be read as vessel..., with the last two characters marked unreadable in the transcript; the black version writes the same sentence as vessel full scale, and both mean the same, so this page uses vessel full scale throughout.
Silver version HY-IND232 (soft keys Fn / SP / CAL0, with the zero key)
- Enter the program: press the zero key and fn at the same time → the indicator shows F1 → press SP to confirm
- Select the menu:
CAL0selected with the FN key → press SP to confirm - Enter the password: the FN key makes the indicator flash E2 and then show 0 → enter the password (zero key increases the value, FN key shifts) → press SP to confirm
- Division: the display shows
ddivision → select the division with FN → press SP to confirm - Full scale:
Cfull-scale setting → FN key shifts, zero key increases the value → press SP after setting - Empty-scale check: the indicator shows
noLAd→ empty-scale check → press SP to confirm - Load test weights: after the empty-scale check the indicator shows
AddLd1→ load the test weights → press SP to confirm - Enter the test weight: FN key shifts, the zero key increases the value → enter the weight of the test weights loaded → press SP after entering
- Second linearity: the indicator shows
AddLd 2; the second linearity needs no calibration - Exit: press the zero key to exit (caption: calibration complete)

Silver version: CAL0 selected with the FN key (video at 00:13). The soft-key path differs from the other two versions and is the largest difference of the three.
| Item | Black HY-IND232 | Blue HY-IND232+ | Silver HY-IND232 |
|---|---|---|---|
| Enter the program | Press 0 and fn together | Press FN and the zero key together | Press the zero key and fn together |
| Key type | Physical keys | Physical keys | Soft keys |
| Confirm key | ENTER | Confirm key | SP |
| Shift / increase | T shifts, 0 increases | T shifts, 0 increases | FN shifts, zero key increases |
| Level numbers | F1.1 / F1.2 / F1.3 | FN1.1 / FN1.2 / FN1.3 | d → C → noLAd → AddLd1 |
| Level for loading test weights | F1.3 load test weights | FN1.3 load test weights here | AddLd1 load test weights |
| Closing caption | Calibration complete | Returns to 0, calibration complete | Calibration complete |
What are the two key parameters?
These two sentences are written directly on the silver-version video captions, reproduced here:
-
Division = (module capacity × number of modules) ÷ 3000
From the company training video (silver version, 00:34.5-00:36.0).

Original frame, uncropped: the screen shows the
ddivision screen, and the caption is that formula. -
Test weight = 10% to 20% of the total of the modules
From the company training video (silver version, 01:27.0-01:30.0).

Original frame, uncropped: the screen shows the
AddLd1loading screen, and the caption is that sentence.
Both sentences are written on the video frames and describe how the value is chosen for this indicator; the actual value still has to be set from the module configuration and the conditions on site.
What should I check during the procedure?
- Press the confirm key after the test weights are loaded, and after the change - confirm every value you change.
- After the test weights are removed the indicator returns to zero. The order in the video is: the indicator shows 0.5 kg → remove the test weights → the indicator returns to zero automatically.
- What the full scale is based on: the silver-version caption reads: the full scale is set from the maximum load of the equipment or vessel on site.
- The silver version checks the empty scale first: the indicator shows
noLAd→ empty-scale check → press SP to confirm; only after that doesAddLd1appear for loading test weights. - The second linearity is not calibrated: the silver-version caption reads: the indicator
AddLd 2second linearity needs no calibration. - How the silver version exits: press the zero key.
- Do not copy the values from the video: the captions themselves say we change it according to our own needs and modify it according to our own needs - division, full scale and test weight all depend on the site.
To check a step against the original clip, or if you are stuck on site:
Still unsure which version you have?
Tell us the model on the nameplate, the keys on the front, and what the screen shows when you start.