What is a weighing module, and where does it mount?
Short answer: a weighing module is a structural accessory for a load cell, made of a top plate, a base plate, the load cell, a load-bearing column and support bolts. (Public technical literature, "Selection and installation of weighing modules".)
On this model the physical form is: a stainless mounting plate with mounting holes at the bottom; a column-type load cell with a stainless hermetically sealed housing in the middle; a top plate above; and on the side a tie rod passing through an angle-steel bracket that holds the upper and lower parts together, limits lateral movement and prevents tipping. The photographs and the assembly drawing below show this form.
The module mounts under the weighing vessel (feet, pallets) or on its upper part (the waist of a reactor, tank or hopper); all modules of one weighing system must sit in the same horizontal plane. (Public technical literature, "Selection and installation of weighing modules".)
What changes with vessel size is not whether a module is needed, but that lateral force and tipping risk are larger: wind load, agitation, and the shock of filling and discharge all act on the vessel. That is why this model carries a tie rod and an angle-steel bracket as side restraint, and why the company documentation lists the safety support block marking for overload protection among the main features.
Where does it sit in a weighing system?
A weighing control system is normally made of a weight indicator, a junction box, the weighing modules and cable: the modules carry the load, the junction box sums the signals of several modules, and the indicator displays and controls. (Company installation instructions.)
Three to six modules, up to twelve, make up one weighing system. The foundation carrying the modules must meet the stiffness requirement and lie in one plane, or be adjustable to one plane, so that every module carries an even share. (Company installation instructions.)
What is the difference between HY-GW and HY-GWC?
Short answer: the company documentation is identical word for word on main features, wiring and capacity for HY-GW and HY-GWC, and states no difference between them. Order by the exact model and confirm against the shipping documentation; do not infer a difference from the suffix.
| Item | HY-GW (this model, company documentation) | HY-GWC (company documentation) |
|---|---|---|
| Main features (7 items) | Large capacity, low dead weight; safety support block marking for overload protection; stainless-steel hermetically sealed load cell at IP68, usable in harsh environments; simple and quick installation; easy maintenance, less downtime; intended for batching control on large tanks in harsh environments; stainless mounting plate Company documentation, transcribed | Identical word for word Company documentation |
| Wiring (6 wires) | Green = +excitation; black = −excitation; white = +signal; red = −signal; yellow = +sense; blue = −sense Company documentation, transcribed | Identical word for word Company documentation |
| Capacity (5 steps) | 15, 20, 30, 50, 100 t Company documentation, transcribed | Identical word for word Company documentation |
The table compares these three items only. Both right-hand cells read "identical" because the two documents carry the same text for them — nothing is left out here.
Why does this model have a page of its own, then?
- This model has photographs under its own model number: they show this body, and an image of the other model cannot stand in for it.
- For two models that differ only in a suffix, the first question is what differs: this page answers it and gives an action to take — check the model and the shipping documentation — instead of leaving it to guesswork.
- No invented difference: the company documentation has no wording such as "the later version" or "the extra structure", so none is written and none is inferred here.
Does HY-GWC have any images, and why is there no separate page for it?
Short answer: the company documentation provides no images at all for HY-GWC — no photograph, no assembly drawing, no wiring diagram, no dimensioned view. Its main features, wiring and capacity are identical word for word to HY-GW, so this site has no separate page for it: the side-by-side table and the specification table on this page apply to HY-GWC as well.
The reason is to avoid two pages with almost the same content: one model family, one piece of company documentation, split over two pages only makes the question "what is HY-GWC" harder to answer. When photographs or drawings for HY-GWC arrive, a page of its own can follow.
No photographs or drawings of HY-GWC are published here (none were provided). Orders, enquiries and shipping documentation are still handled by model, so quote the full suffix.
What are the specifications, and where does each value come from?
Short answer: the table lists the documented values one by one with their source, and the right-hand column translates each value into what it decides at selection and installation. Values are transcribed as given: no conversion, no merging, and no steps or items that the documentation does not contain.
| Parameter | Value (source) | What it determines |
|---|---|---|
| Model | HY-GW Company documentation | Ordering and enquiry reference. The related model HY-GWC shares the same wording on features, wiring and capacity in the company documentation, and no difference is stated, so always quote the full suffix. |
| Category | Weighing module Company documentation | A structural part plus a load cell: capacity, material and mounting dimensions all matter together, not the cell alone. |
| Capacity steps | 15, 20, 30, 50, 100 t Company documentation (5 steps, transcribed) | The load limit per module. Check dead load plus live load against the rated capacity of one module and take the next step up. The largest vessel does not automatically take the largest step; the step follows the calculated load per module. |
| Material and sealing | Stainless-steel hermetically sealed load cell; stainless mounting plate Company documentation | Where it can be used: a welded stainless housing plus a stainless mounting plate is the basis for damp, washdown and corrosive-atmosphere shops. For aggressive media or high temperature the documentation gives no rating, so check against the actual conditions. |
| Protection | IP68 Company documentation | Applies to the sealed load cell body. Large vessels often stand outdoors or are washed down, so check the cable entry and the junction box sealing to the same standard: replacing the module alone does not stop water entering the junction box. |
| Structural features | Large capacity, low dead weight; safety support block marking for overload protection; simple and quick installation; easy maintenance, less downtime Company documentation | Large capacity at low dead weight means less dead load added to the foundation at the same tonnage. The support block marking is the datum for overload protection: confirm the support position against it at installation and service, and do not move it. After mounting, check verticality and both mounting faces against the table below. |
| Wiring system | 6-wire: excitation +/−, signal +/−, sense +/− Company documentation | The two sense wires compensate the excitation loss over a long cable run, which is what makes them useful when the cable from the module to the indicator or junction box is long. Whether to use them, and how to terminate them, follows the shipping documentation and the wiring diagram of the unit. (Function from public technical literature.) |
| Colour code | Green = +excitation; black = −excitation; white = +signal; red = −signal; yellow = +sense; blue = −sense (6 wires) Company documentation | Field wiring reference. The nameplate on the unit in the photographs shows a second code (+EXC red, −EXC black, +SIG green, −SIG white, SHIELD thick yellow), the same one as in the company installation instructions. Both are reproduced here and neither is chosen for the customer. Before wiring, follow the shipping documentation and the unit's own wiring diagram; do not copy colours across batches. |
| Documented applications | Batching control on large tanks in harsh environments Company documentation, transcribed | A direction — harsh environment, large tank — not a promise that any large vessel fits. Check vessel weight, product weight, number of feet and site conditions for each case. |
| External dimensions | Values marked on assembly drawing HZS01-10~50t (see the dimensions section below) Company drawing | Foundation layout and where the feet land. A different capacity step changes the outline and the hole pattern, so fix the step before a foundation is poured or feet are reworked, then check the hole centres against the drawing. |
Ten rows. Values and wording are transcribed from the company documentation; every explanation added here carries its own source and comes from public technical literature rather than from the company. Accuracy, long-term stability, material grade, temperature range and corrosion resistance are not stated in the documentation for this model, so they are not filled in and not inferred here.
Why are there two wiring colour codes?
Short answer: the product documentation text, reproduced in the table above, gives the set with green = +excitation, 6 wires: green = +excitation, black = −excitation, white = +signal, red = −signal, yellow = +sense, blue = −sense. The nameplate on the load cell of the unit in the photographs gives another set: +EXC (red) positive excitation, −EXC (black) negative excitation, +SIG (green) positive signal, −SIG (white) negative signal, SHIELD (thick yellow) shield. That set matches the company installation instructions, where red = excitation +.
The colour-to-signal mapping is reversed between the two: the same green wire is +excitation in the product text and +signal on the nameplate. Colours must therefore not be copied across batches. Before wiring, follow the shipping documentation and the wiring diagram of the unit: whichever set the nameplate of the unit in hand shows, wire to that set.
This section reports what the documentation and the photographs contain. It does not judge which set is newer or which one is right; that call needs a physical check on the unit and is not replaced here.
What are the external dimensions, and which capacity steps does the drawing cover?
Short answer: one assembly drawing is filed for this model: the HZS01-10~50t module assembly drawing (drawing number IN HZSO1-10~50t-M-00). The values marked on it are transcribed below. The drawing marks no units, and none are added here.
| View | Values marked on the drawing | Source |
|---|---|---|
| End view (K direction) | Length 280, segments 25 / 115 / 115; width 200, segments 50 / 88 / 16.5 / 38; upper 196 / 185 / 160 / 82.5; hole groups 6-Φ18, 9-Φ18 | Company drawing |
| Side view (section) | Overall height 222.5; top plate 15; base plate 15; right-hand vertical 65 / 102.5 | Company drawing |
The title block reads "HZS01-10~50t module assembly drawing" and the parts list entry for the load cell reads "10~50t load cell", so the drawing covers the 10–50 t range. This model's capacity range includes the 100 t step, and the company documentation does not say whether 100 t uses the same drawing, so nothing is inferred here: check the dimensions of the 100 t step against its own shipping documentation. Two further points: the drawing marks no units (every value is transcribed without adding mm), and it does not label which capacity step each dimension belongs to.
Which values apply to which working conditions?
Short answer: size the capacity with dead load + live load ≤ rated capacity × number of modules × 70%, then take the next step up; take one module per actual support point. On a large vessel, look at lateral force and foundation stiffness as well. (Criteria from public technical literature.)
dead load + live load ≤ rated capacity × number of modules × 70%
The formula is repeated as plain text so it can be copied, sent to a work group or printed. Worked examples are on the Selection & Resources page.
| Your case | How to take it | Basis |
|---|---|---|
| The vessel already has support feet | Set the module count by the number of support feet: three feet, three modules; four feet, four modules | Public technical literature |
| New vessel, vertical cylinder | Three-point support is stable; where wind, sway and vibration act on it, four-point support is the better choice | Public technical literature |
| The calculated load per module falls between two steps | Take the next step up, never the step below | Public technical literature |
| How many to buy for one set, and how to mix the structures | One fixed and one semi-floating module per system, the rest floating; sets are usually three or four modules | Public technical literature |
| The vessel stands outdoors and sees wind load | The foundation needs enough bearing capacity and stiffness (deflection below 3 mm); skew of the upper bracket and of the base bracket from horizontal not more than 0.5° each; this model carries a tie rod and an angle-steel bracket as side restraint, to be assembled as the drawing shows | Foundation and skew from public technical literature; side restraint per the assembly drawing |
| The site is damp and gets washed down | This model has a stainless hermetically sealed load cell at IP68 and a stainless mounting plate; check the cable entry and the junction box with the same sealing in mind | Values from company documentation |
| Replacing an old module without changing the foundation | Check the hole pattern and the overall height against the assembly drawing first (values in the section above), then against the shipping documentation and the unit; do not infer the 100 t dimensions from the 10–50 t drawing | Company drawing; no drawing for the 100 t step |
Where does it fit, and where does it not?
Short answer: what the company documentation and the sourced content on this page support is batching control on large tanks in harsh environments, plus the damp, washdown and corrosive-atmosphere conditions that go with a stainless hermetically sealed cell at IP68 and a stainless mounting plate. Items the documentation does not give — accuracy and long-term stability, material grade, temperature range, corrosion resistance, the 100 t dimensions — are not inferred here.
Where it fits (documented)
- Large tanks in harsh environments: the documented wording is "intended for batching control on large tanks in harsh environments". (Company documentation.)
- Outdoor vessels and vessels under wind load: this model carries a tie rod and an angle-steel bracket as side restraint, and the assembly drawing marks the restraint screw and the spherical bearing; assemble them as drawn. (Company drawing.)
- Where an overload protection datum is needed: the documentation states a safety support block marking for overload protection. (Company documentation.)
- Damp shops and washdown areas: based on the stainless hermetically sealed load cell, IP68 and the stainless mounting plate; the cable entry and the junction box must be sealed to the same standard. (Values from company documentation.)
- Turning a vessel into a weighing system: whether tank, hopper or reactor, fitting weighing modules and an indicator turns it into a weighing system. (Public technical literature, "Selection and installation of weighing modules".)
Boundaries and limits (read these too)
- Accuracy, long-term stability, material grade, temperature range and corrosion resistance are not given: these are not inferred here and no assumed value is published; check them against the process and the site. (The company documentation does not record them.)
- No drawing for the 100 t step: the one drawing covers 10–50 t, so the 100 t dimensions cannot be taken from it. (Drawing title and parts list say 10–50 t.)
- Do not infer from the suffix what HY-GW has more or less of than HY-GWC: the company documentation is identical word for word on the three items and states no difference. (Company documentation.)
- Never mix capacities within one weighing system: check the capacity of every module before mounting. (Company installation instructions.)
- No welding current through the load cell: no other current, welding included, may pass through the cell. (Company installation instructions.)
- The vessel must not be mechanically tied to parts the modules do not carry, or the reading will be wrong. (Public technical literature, "Selection and installation of weighing modules".)
- Pipes left as they are will spoil calibration: run every pipe connected to the vessel horizontally so it does not hang the vessel, and use flexible connections. (Public technical literature, "Selection and installation of weighing modules".)
What does it look like, and is the drawing legible?
Short answer: the company documentation holds 6 images: 2 photographs (used here), 3 not used (the markings and model information on those nameplates are not for publication), and 1 assembly drawing (used here, for the dimensions). Images are local files; the page loads no external resource.
All three images are used unaltered and carry a faint Hengyuan watermark. Three further photographs are not used: the load cell or base nameplate on them carries markings and model information that are not for publication, so they are neither shown nor referenced. Where an image and the written documentation disagree, the written documentation governs.
How is it installed and wired?
Short answer: three things decide whether a module sits right — the stiffness and level of the base, the verticality of the module against its mounting face, and whether outside forces such as pipes and welding have been taken out of the load path. Each row below gives a value or an action to check. Wire to the shipping documentation and the wiring diagram of the unit; the colour code follows the nameplate of the unit in hand.
| Item | Requirement | Source |
|---|---|---|
| Verticality | The vertical axis of every module must line up with the perpendicular to the mounting base plate, deviation not more than 1.5°; where the face is uneven or out of square, correct it with shims | Company installation instructions |
| Frame and base stiffness | The frame and base plate carrying the modules must be stiff enough; stiffness must reach 1/1000 | Company installation instructions |
| Mounting faces | Top and bottom mounting faces of every module fully in contact, flat throughout, and evenly loaded | Company installation instructions |
| Levelling | After mounting three or four modules under the vessel feet or brackets, rock the top plate by hand; if it moves, the module is not level — shim it until it no longer moves | Company installation instructions |
| Bolting | Tighten all bolts diagonally: pre-tighten first, then torque to specification | Company installation instructions |
| Fixing method | Once the module is level, either bolt the top plate to the vessel feet or brackets and the base plate to the steelwork or floor, or weld both (note: during welding no current may pass through the module) | Company installation instructions |
| Welding | No other current, welding included, may pass through the load cell; it will damage the cell | Company installation instructions |
| Cable | Do not pull the cable tight, so it is not under tension and its electrical properties are not affected; protect the cable from cuts and abrasion | Company installation instructions |
| Handling | Do not lift the module by its cable when taking it out of the packaging; avoid knocks and never let it fall from a height | Company installation instructions |
| Module capacity | Check the capacity before mounting; never mix capacities within one weighing system | Company installation instructions |
| Wiring and check | Connect the load cell cables to the junction box as the wiring diagram shows, complete the run from the junction box to the indicator, re-check every connection, and verify the supply voltage | Company installation instructions |
| Bracket skew | Skew of the upper bracket from horizontal not more than 0.5°; twist or skew of the base bracket from horizontal not more than 0.5° | Public technical literature |
| Foundation | The foundation or floor carrying the vessel must have enough bearing capacity and stiffness; deflection below 3 mm; mounting on a concrete foundation is the better choice | Public technical literature |
| Piping | Run every pipe connected to the vessel horizontally so it does not hang the vessel; put the first pipe support as far from the vessel as possible; use flexible connections | Public technical literature |
| Independent load path | The vessel must act on the load cells alone and have no mechanical tie to parts the cells do not carry | Public technical literature |
| Earthing | One independent earth per module | Public technical literature |
| Service access | Mount the modules where they can be reached for service | Public technical literature |
| Calibration conditions | Allow for test-weight positions and jacking points at the design stage of the vessel; calibrate after installation | Public technical literature |
The rows above are the general requirements of the company installation instructions and of public technical literature for this class of module; site practice and the final decision still follow the shipping documentation and the actual conditions. No wiring diagram for this model was provided, so wiring follows the shipping documentation and the unit's own diagram; the tie rod and angle-steel bracket are assembled as the assembly drawing shows.
Where do the facts come from?
- Company product documentation (HY-GW product page text): model, category, capacity steps, large capacity at low dead weight, the safety support block marking, the stainless hermetically sealed load cell and IP68, the stainless mounting plate, 6-wire connection and colour code, documented applications; also the material for the word-for-word check of HY-GW against HY-GWC.
- Company drawing (HZS01-10~50t module assembly drawing, number IN HZSO1-10~50t-M-00): outline dimensions, mounting hole groups, tie rod and restraint parts, parts list.
- Company installation instructions: what a module and a weighing system consist of, module count and foundation requirements, mounting tolerances and stiffness, the fixing and welding prohibitions, cable and wiring requirements, and the second wiring definition (red = excitation +).
- Two photographs of this model: the form of the unit and the wiring code on its nameplate.
- Public technical literature ("Selection and installation of weighing modules" and similar): module construction, sizing and module count, piping and calibration requirements, static and dynamic criteria.
Values are transcribed from company documentation and are not a commitment for a specific project; final specifications are confirmed against the application. Dimensions follow the assembly drawing, which covers 10–50 t and marks no units; the documentation provides no drawing for the 100 t step.