First published · Chinese version: HY-UWS 称重模块

You are here: Home > Products > HY-UWS

HY-UWS Weighing Module: 5 kg to 500 kg, 10 Steps, Stainless Steel, IP68

HY-UWS is a stainless-steel weighing module: load cell, top plate, base plate, support column and support bolts in one unit, mounted under the feet or brackets of a tank, silo or reactor. It turns weight into an electrical signal for the weight indicator and is used for batching control.

If you cannot tell which capacity step fits, describe the duty:

+86 183 5123 8137

One tap dials the number — no form to fill in.

Short answer: HY-UWS covers 5, 10, 20, 30, 50, 75, 100, 200, 300 and 500 kg in 10 steps; it uses a stainless-steel hermetically sealed load cell with IP68 protection; it is a 6-wire module with a spherical joint and support bolts. All values are transcribed from the company product documentation, with the source of every row given. Three things have to be said up front: (1) the company documentation for HY-UWS, HY-UW, HY-UW-M and HY-UWS-M is word-for-word identical on three items and states no difference between them (see the section below); (2) the nameplates on the units photographed for this page show two different wiring codes; both are reproduced here and the page does not choose one for the customer (see the note under the parameter table); (3) no drawing was supplied for this model, so this page publishes neither a wiring diagram nor dimensions.

Values are transcribed from company product documentation; every field carries its source in the table.

This model belongs to the HY-UW series — four models documented with the same set of parameters, with no stated difference between them: all models, grouped by capacity

What is a weighing module and where does it mount?

Short answer: a weighing module is a structural accessory for a load cell, made of a top plate, a base plate, the load cell, a load-bearing column and support bolts. (Structure from public technical literature on weighing module selection and installation.)

On this model the physical form is: a top plate with mounting holes on top, a metal bellows body in the middle (the housing of the hermetically sealed load cell), a base plate with mounting holes below, and a cable leaving from the side. Support bolts sit between top plate and base, and a braided copper strap bonds the top plate to the base. (Physical form from the photographs below; structural description from the company installation instructions.)

A module usually mounts at the bottom of the weighing structure — feet, platform, and so on — but may also mount on the upper part of the structure, on the waist of a vessel, tank or hopper. All modules of one weighing system have to sit in the same horizontal plane. (Public technical literature on weighing module selection and installation.)

That is the difference from buying a load cell on its own: a module is the load cell, the load-bearing structure and the mounting connections built into one part, so it carries load as soon as it is fitted to the feet. A bare load cell leaves the load path, the restraints and the mounting face to the installer. (Public technical literature on weighing module selection and installation.)

Where it sits in a weighing system

A weighing control system is usually made of the display and control indicator, the junction box, the weighing modules and the cable. The modules carry the load, the junction box collects the signals of several modules, and the indicator displays and controls. (Company installation instructions.)

Modules connect easily to structures of any shape — conveyors, platforms, vertical tanks, vessels, hoppers — to build a weighing system for a given duty. A system usually uses 3 to 6 modules, and no more than 12, as one set. The foundation carrying the modules has to meet the stiffness requirement and lie in one horizontal plane, or be adjustable to it, so that every module carries its share evenly. (Company installation instructions.)

What is the difference between HY-UWS and HY-UW, HY-UW-M and HY-UWS-M?

Short answer: for HY-UWS, HY-UW, HY-UW-M and HY-UWS-M the company documentation is word-for-word identical on main features, wiring and capacity, and states no difference between the four. Order by the exact model and confirm against the shipping documentation of the unit; do not read a difference into a suffix.

The table below compares those three items. The left column names the item, the second column carries the documented wording for this model, and the three columns on the right give the result for the other models in the family.

HY-UWS against HY-UW, HY-UW-M and HY-UWS-M: the three items, word for word
Item HY-UWS (documented text) HY-UW HY-UW-M HY-UWS-M
Main features (6 items) Stainless-steel hermetically sealed load cell with IP68, for use in harsh environments; spherical joint keeping the module vertical under load; support bolts against tipping; simple and quick installation; easy maintenance, saving downtime; for batching control on small and medium reactors in chemical, pharmaceutical and food plantsCompany documentation (transcribed) Identical Identical Identical
Wiring (6 wires) Green = +excitation; black = −excitation; white = +signal; red = −signal; yellow = +sense; blue = −senseCompany documentation (transcribed) Identical Identical Identical
Capacity (10 steps) 5, 10, 20, 30, 50, 75, 100, 200, 300, 500 kgCompany documentation (transcribed) Identical Identical Identical

Only these three items are compared. Every column reads identical because the four documents carry the same wording on these items — nothing has been left out on this page.

So why does this model have its own page?

What are the HY-UWS specifications, and where does each value come from?

Short answer: the table lists the documented values of HY-UWS row by row with the source, and the right-hand column turns each value into the decision it drives at selection and installation. Values are transcribed as they stand — not converted, not merged, and no capacity step or item is added that the documentation does not contain. No drawing was supplied for this model, so the table has no "wiring diagram" row and no "dimensions" row (the reason is at the end of this section).

HY-UWS specifications, sources and what each value determines
Parameter Value (source) What it determines
Model HY-UWS Company documentation The identifier used when ordering. On the three shared items this model and HY-UW, HY-UW-M and HY-UWS-M carry word-for-word identical documented wording with no difference stated (see the section above), so quote the full suffix and check it against the shipping documentation of the unit.
Category Weighing module Company documentation It is a structural part plus a load cell, so capacity, material and mounting dimensions are checked together — comparing load cell figures alone is not enough.
Capacity steps 5, 10, 20, 30, 50, 75, 100, 200, 300, 500 kg Company documentation (10 steps, transcribed) The maximum load one module carries. Total capacity of a system = rated capacity × number of modules, so calculate with vessel dead weight and product weight first, then take the next step up.
Material and sealing Stainless-steel hermetically sealed load cell Company documentation Sets the environments it can work in: a welded stainless housing is one precondition for damp areas that need washdown. For aggressive media or high temperature the documentation gives no rating — check against the duty.
Protection IP68 Company documentation The documented rating applies to the sealed module body. On site the cable entries and the junction box have to be sealed to the same standard — a new module alone does not fix water getting into a junction box.
Structural features Spherical joint keeping the module vertical under load; support bolts against tipping; simple and quick installation; easy maintenance, saving downtime Company documentation The spherical joint absorbs small angular errors of the mounting face and keeps the module in a vertical weighing position; the support bolts decide whether the structure can tip. Both are checked after mounting, against the verticality and mounting-face rows of the installation table below.
Wiring system 6-wire (excitation + / −, signal + / −, sense + / −) Company documentation Six wires add +sense and −sense to the four of a 4-wire cell: by public technical literature these two carry the compensation of excitation loss over a long cable run, which is why they matter when the cable from module to indicator (or junction box) is long. Whether they are used, and how, follows the shipping documentation and the wiring diagram of the unit.
Colour code Green = +excitation; black = −excitation; white = +signal; red = −signal; yellow = +sense; blue = −sense (6 wires) Company documentation The field wiring reference — but two codes appear under this model number: one written by colour (the row above, with an extra "yellow (long) = shield" line on one nameplate), one written by English terminal name (+EXC in red, +SIG in green), which is the code in the company installation instructions. Both are reproduced below and neither is chosen for the customer. Before wiring, follow the shipping documentation and the wiring diagram of the unit; do not copy colours across batches.
Documented applications Batching control on small and medium reactors in chemical, pharmaceutical and food plants Company documentation (transcribed) An industry direction, not a claim that every plant in those industries fits. Check vessel dead weight, product weight, number of feet and site conditions for each case.

Values and wording are transcribed from company documentation; every explanation added on this page carries its own source and comes from public technical literature, not from a company commitment. Nine rows.

Why are there two wiring colour codes?

Short answer: the main text and the table above reproduce the company product page set, six wires: green = +excitation, black = −excitation, white = +signal, red = −signal, yellow = +sense, blue = −sense. Two other versions are visible in the same documentation set under this model number; both are reproduced here, neither is merged with the other, and the page does not pick one for the customer:

In short: the product page text and the nameplate on photo 1 belong to the "green = +excitation" set; the nameplates on photos 2 and 4 and the company installation instructions belong to the "red = EXC+" set. Both have appeared under this model number and neither cancels the other. Before wiring, follow the shipping documentation and the wiring diagram of the unit; do not copy colours across batches — wire to the nameplate of the unit you have.

This section states only what the documents and the photographs contain. It does not judge which set is newer or which set is right; that needs a check against the physical unit and is not replaced by this page.

Why are these rows missing?

Short answer: the company documentation states no combined accuracy or long-term stability figure for this model and provides no drawing for it, so the wiring diagram and the dimensions are left to the shipping documentation. No value is inferred here.

The drawings are a separate case from a missing number: no drawing was supplied for this model, so the wiring diagram and dimension rows are left out entirely rather than filled with a "not stated" line. Nothing is borrowed from the other models in the family either — their drawings carry their own dimension steps, and a borrowed one would be a wrong value. Use the shipping documentation of the unit for both.

Which capacity step fits a given duty?

Short answer: check the load per module against "dead load + live load ≤ rated capacity × number of modules × 70%" and take the next step up from the result; take one module per actual support point. (Criterion from public technical literature.)

dead load + live load ≤ rated capacity × number of modules × 70%

The same formula in words, so it can be copied, sent to a work group or printed. Worked examples are on the Selection & Resources page.

Case by case (follow directly)

HY-UWS: which capacity step and how many modules
Your case What to do Basis
The vessel already has support feet Take the number of modules from the number of feet: 3 feet, 3 modules; 4 feet, 4 modules Public technical literature
A new vertical cylindrical vessel Three-point support is stable; where wind, sway or vibration act on the vessel, four-point support is the better choice Public technical literature
The calculated load per module falls between two steps Take the next step up, never down Public technical literature
How many modules and which types in one set One fixed module and one semi-floating module per set, the rest floating; sets are usually three or four modules Public technical literature
A damp area that needs washdown This model is a stainless-steel hermetically sealed load cell with IP68; check cable entries and the junction box against the same sealing requirement Values from company product documentation
Six wires to terminate, one pair of sense wires Terminate each of the six positions at the junction box and the indicator; see the colour code above and use the nameplate set of the unit in hand Wire functions from company product documentation; practice from company installation instructions
Replacing an old module without rebuilding the foundation No drawing is supplied for this model, so no dimensions are given here: measure the existing hole pattern and the available height first, then check against the shipping documentation and the unit Company documentation provides no drawing for this model

Where does it fit, and where does it not?

Short answer: what can be supported by the company product documentation and the sourced statements on this page is batching control on small and medium reactors in chemical, pharmaceutical and food plants, plus the damp and washdown environments implied by a hermetically sealed stainless cell with IP68. Items the documentation does not give — combined accuracy and long-term stability, material grade, temperature range, corrosion class, a drawing for this model — are neither inferred nor decided for the customer here.

Fits, where documented

Limits and exclusions, equally important

What does it look like, and are drawings supplied?

Short answer: the three photographs below are of this model — the unit itself, another angle, and the nameplate on the side of the load cell. This model has no drawing: the company documentation supplies no wiring diagram and no dimension drawing for it, and drawings of other models in the family are not used as a substitute. The images are local files; the page loads nothing from outside.

HY-UWS weighing module: top plate with mounting holes, metal bellows body, base plate with mounting holes, cable leaving from the side, support bolt and a braided copper strap, with a wiring nameplate on the side of the load cell
Photo 1 — front view of the unit: top plate, metal bellows body, base plate and the cable leaving from the side. The nameplate on the side of the load cell writes the wiring by colour in Chinese, matching the product page text, with an extra line "yellow (long) = shield". Company documentation; the original carries a light Hengyuan watermark.
HY-UWS weighing module from another angle: top plate mounting holes, metal bellows body, support bolt and nut, base plate mounting holes, braided copper strap and side cable, with a nameplate written by English terminal names
Photo 2 — another angle: top plate holes, bellows body, support bolt and nut, base plate holes, braided copper strap and side cable. The nameplate on this unit is written by English terminal name (+EXC (red), +SIG (green) and so on). Company documentation; the original carries a light Hengyuan watermark.
HY-UWS weighing module, side view of the load cell: metal bellows body, nameplate with the wiring written by English terminal names, support bolt and base plate
Photo 3 — side view of the load cell: bellows body and the wiring nameplate, written by English terminal name (+EXC (red), −EXC (black), +SIG (green), −SIG (white), shield). The wording on this page is not rewritten from the image. Company documentation; the original carries a light Hengyuan watermark.

All three photographs are company product images, used unaltered, each carrying a light Hengyuan watermark. The company documentation supplies no drawing for this model, and no drawing of another model has been used to fill the gap, so this section has no wiring diagram and no dimension drawing. Where an image and the written documentation disagree, the written documentation governs.

How is it installed and wired?

Short answer: whether a module installs well comes down to three things — stiffness and level of the base, verticality of the module against its mounting faces, and whether external paths such as pipework and welding have been removed. Each row below gives a value or an action that can be checked. Wire to the shipping documentation and the wiring diagram of the unit; the colour code follows the unit in hand.

Installation and wiring requirements, source by source
Item Requirement Source
Verticality The vertical axis of every module must coincide with the perpendicular to the mounting plate, deviation no more than 1.5°; correct any unevenness or lean with shims or equivalent Company installation instructions
Frame and base stiffness The frame and base plate carrying the modules must be stiff enough; stiffness has to reach 1/1000 Company installation instructions
Top and bottom mounting faces Top and bottom faces of every module fully connected, with no unevenness, and evenly loaded Company installation instructions
Levelling After mounting 3 or 4 modules under the vessel feet or brackets, rock the top plate by hand; if it moves, the module is not level — shim with steel plate until it no longer moves Company installation instructions
Bolt tightening Tighten all bolts diagonally: pre-tighten first, then torque to specification Company installation instructions
Fixing method Once a module is level, either bolt it — top plate to the vessel foot or bracket, base plate to the steelwork or floor — or weld it (note: no welding current may pass through the module) Company installation instructions
Welding No other current, welding included, may pass through the load cell, or the cell is damaged Company installation instructions
Cable Do not pull the cable tight, so no tension reaches it and spoils the electrical behaviour; protect the cable from cuts and abrasion Company installation instructions
Handling Never lift the module by its cable when taking it out of the packing; avoid knocks and do not let it fall from height Company installation instructions
Module capacity Check module capacity before mounting; never use modules of different capacity in one weighing system Company installation instructions
Wiring and re-check Connect the load cell cables to the junction box as the wiring diagram shows; complete the run from junction box to display and control indicator; re-check every connection; check the supply voltage Company installation instructions
Bracket skew Skew of the upper bracket against horizontal, and skew or twist of the base bracket against horizontal, must each stay within 0.5° Public technical literature
Foundation The foundation or floor carrying the vessel must have enough bearing capacity and stiffness, with deflection below 3 mm; a concrete foundation is the better choice Public technical literature
Pipework Run every pipe connected to the vessel horizontally so no pipe hangs the vessel; put the first pipe support as far from the vessel as possible; use flexible connections where possible Public technical literature
Independent load path The vessel must act on the load cells alone, with no mechanical tie to any part the cells do not carry Public technical literature
Earthing Each module gets its own earth Public technical literature
Service access Mount the modules where they can be reached for maintenance Public technical literature
Calibration conditions Allow, at design stage, for test-weight positions and for jacking points used during installation and maintenance; calibrate after installation is complete Public technical literature

The rows above are the general requirements of the company installation instructions and public technical literature, and apply to modules of this kind; the practice on site and the final acceptance are still set by the shipping documentation and the actual conditions. No wiring diagram for this model is supplied by the company, so wiring follows the shipping documentation and the wiring diagram of the unit. Indicator-side calibration steps are covered in the FAQ.

Where do the facts come from?

Values are transcribed from company documentation and do not constitute a commitment for a specific project; specifications are confirmed against the duty. No drawing was supplied for this model, so the wiring diagram and dimensions come from the shipping documentation.

Still unsure which capacity step fits?

Give us these in one go and a capacity step with module count can be suggested directly: vessel dead weight, product weight, number of support feet and their layout, product name and behaviour, site conditions, and whether it is a new build or a retrofit. Quote the full model suffix when ordering.

Call +86 183 5123 8137