Which fault should I check first?
Short answer: do not work from H1 to H8. Work in order of check cost — start with what needs no tool, no shutdown and a few minutes, and touch the load cell last. Get the order wrong and you circle around inside parts that all look fine.
| Order | Check first | How to check | Cost |
|---|---|---|---|
| 1 | Top nut on the support bolt loosened, sleeve free (H1) | Hand check. No tool needed. | 1 minute, no shutdown |
| 2 | Product pipe: rigid or flexible connection (H4) | Visual; loosen the pipe joint and watch whether the reading changes | 10 minutes, feed may have to stop |
| 3 | Horizontal restraint bolts not bottomed out (added check) | Set the gap from the drawing; without a drawing, push the vessel — slight movement, no rocking | 30 minutes |
| 4 | Indicator supply voltage inside the permitted range (H8) | Multimeter across the supply terminals | 10 minutes |
| 5 | Auto zero on power-up switched off (H2) | Walk the indicator menu and read the state of that setting | 5 minutes |
| 6 | Junction box damp, spare entries plugged (H7) | Open the box: moisture, condensation, gasket hardened | 20 minutes |
| 7 | Whether it is due a calibration (acceptance section) | Load at 20% / 50% / 100% and check repeatability | Half a day, needs test weights |
| 8 | Only now suspect the load cell itself | Measure input/output resistance and compare with a known-good module of the same model | Needs a meter, may need removal |
This table is the core of the page. The eight faults below are numbered H1 to H8, but check them in the order above.
What does each of the eight faults look like, how do I confirm it and how do I fix it?
H1 Nothing weighs at all after installation
On site you see: you put a load on and the indicator barely moves, or it sits at 0.
Cause: the nut on top of the support bolt was never loosened, so the sleeve is not free. The module is propped up and the weight never reaches the load cell.
How to confirm: look at the support bolt on the module — if the top nut is still tight, that is it. Loosen it and push the top plate by hand; it should move slightly.
How to fix: loosen the top nut, let the sleeve return to free movement, read again. This is a mandatory step before calibration; modules are usually locked for transport.
Cost: 1 minute, no shutdown.
H2 After one power cut the vessel has "lost" material
On site you see: after a power failure or an indicator restart the vessel clearly still holds product, but the display has dropped by a large amount (or gone to zero).
Cause: auto zero on power-up is enabled. On a vessel scale the indicator takes the product weight currently in the vessel as tare and zeroes it away.
How to confirm: open the indicator menu and read whether that setting is enabled or disabled.
How to fix: set it to disabled. On a vessel scale this is not optional.
Cost: 5 minutes.
H3 What should the full-scale value actually be?
On site you see: an argument about what to enter as full scale for a 5 t vessel. One person works from the vessel, another from the product.
Cause: two separate things are being treated as one —
- the full-scale value in the indicator: a number you enter, used for display and alarms;
- the rated capacity of the module: the capability of the module itself, fixed when you buy it.
How to confirm: read rated capacity from the module nameplate. (Some nameplates also carry markings that are outside the scope of this page; do not infer other parameters from them.)
How to fix: when you check rated capacity, the dead weight of the vessel must be included. Check against the industry criterion:
dead load + live load ≤ rated capacity × number of modules × 70%
Worked examples and the module-count rule are on Selection & Resources.
Cost: 10 minutes to run the numbers; a wrong module means replacement, which is expensive.
H4 How is it still off after calibrating again and again?
On site you see: repeated calibration, reading still off; different test weights, different indicator, same result.
Cause: the product pipe is a rigid pipe. A rigid pipe carries part of the weight straight into the building structure, so the load cells only see the remainder.
How to confirm: loosen the pipe flange (or cut out a short section) and watch the reading. If it changes, that is the cause.
How to fix: replace it with a flexible connection, or lengthen the connecting pipe as far as practical (length reduces the stiffness effect). All pipes connected to the vessel should run horizontally, and the first support should be as far from the vessel as possible.
Cost: half a day to a day of pipe work, with shutdown.
H5 No place for test weights was left on the vessel — how do I calibrate now?
On site you see: calibration is due, and there is nowhere on the vessel to stand the test weights.
Cause: nothing was allowed for at design stage. A vessel rarely has a natural place to put weights.
How to confirm: check the drawing or the site — is there a platform or bracket under the vessel that can carry a load?
How to fix: on a new vessel, weld a test-weight platform at design stage (usually under the vessel, so weights can be loaded and removed). On an existing vessel, assess whether one can be welded on later — before any welding, confirm that the welding current cannot pass through the load cell (see H6).
Cost: zero on a new vessel (added at design); on an existing vessel, shutdown plus hot work, which is expensive.
H6 Can welding on site destroy a load cell?
On site you see: the mounting bracket is welded, and the next day the scale is out and no amount of calibration brings it back.
Cause: welding current ran through the inside of the load cell. The strain gauges and internal wiring are damaged by the current, and nothing shows from outside.
How to confirm: welding on site plus a reading that will not calibrate back = strong suspicion; then measure the input/output resistance of that module with a multimeter and compare it with a known-good module of the same model.
How to fix: a damaged cell can only be replaced. Prevention is to give the current the shortest path — earth to the scale body when welding the top plate, earth to the foundation when welding the base plate — and never let current cross the load cell. The safer method is to disconnect the load cell from the module electrically before welding.
Cost: prevention = 5 minutes of briefing; a burned cell = module replacement, shutdown and cost.
H7 Why do the digits jump as soon as the rainy season starts?
On site you see: fine in dry weather; once it rains, or in the damp season, the digits start jumping.
Cause: the junction box has taken in moisture; or spare cable entries were left open, so damp air goes straight in.
How to confirm: open the box and look for moisture or condensation; check whether the gasket has hardened; check whether the spare entries are plugged.
How to fix: plug spare entries with sealing plugs; anything mounted outdoors needs a protective box against rain; replace hardened gaskets; run the cable in metallic flexible conduit.
Cost: 20 minutes; if the board inside is already corroded, the junction box has to be replaced.
H8 Is drifting always a failed load cell?
On site you see: the reading creeps up or down, a few kilograms over several minutes.
Cause: do not suspect the load cell first. The usual cause is unstable supply voltage at the indicator, which produces drift once it leaves the permitted range; next come environment (temperature, wind) and moisture in the junction box.
How to confirm: first measure the supply voltage against the indicator's rated range; then log a "no load, 30 minutes" run and see whether the reading moves in steps or creeps smoothly — smooth creep is usually drift (environment, supply, moisture); step changes usually point to structure or wiring.
How to fix: fit a stabilised supply if the voltage is unstable; treat moisture as in H7; if it is long-term drift, recalibrate.
Cost: 10 minutes to half a day.
What other five common faults are there?
| Symptom | Cause | Action |
|---|---|---|
| Load at an angle / side force | Mounting not level, support points at different heights, thermal movement of the vessel | Re-level the feet and add restraints |
| Obstruction in the gap | Dust, product or frost entering the moving gap of the module | Clean the moving gap |
| Electromagnetic interference | Variable-speed drives, power cables running alongside signal cables | Signal cable in metallic conduit, kept away from power cables, one reliable earth point |
| Restraint bottomed out | Gap too small, thermal growth closes it up | Adjust the restraint gap and leave a small amount of travel |
| Base settling / support deformation | Foundation not stiff enough, deformation under long-term load | Strengthen the foundation and calibrate again |
What can be prevented before installation?
Short answer: five of the eight above can be designed out before installation, at almost no cost.
- Allow for a test-weight platform at design stage (H5) — one bracket on a new vessel; welding one onto an existing vessel means shutdown and hot work.
- Make every product pipe a flexible connection (H4) — a rigid pipe is the number one cause of "it will not calibrate".
- Set auto zero on power-up to disabled (H2) — 5 minutes, and a power cut no longer wipes the product weight.
- Brief the welders first (H6) — state "earth to the scale body when welding the top plate, earth to the foundation when welding the base plate" and put it in the work order.
- Plug the spare junction-box entries and fit an outdoor box (H7) — 20 minutes, and the rainy season does not repeat the same fault.
When should I call the manufacturer instead of taking it apart?
- A single module under load shows no change in output at all — possibly an open circuit inside the load cell; confirming it needs instruments.
- Resistance clearly different from a known-good module of the same model — do not keep powering it up, or the damage can spread.
- Suddenly out after welding, and calibration will not bring it back — strongly suggests a damaged load cell; the part has to be replaced.
- Modules read very differently from each other and levelling does not close the gap — one may already be damaged, or two models are mixed in one system.
In these four cases doing it yourself turns a small fault into a large one: opening the housing or forcing the internal structure affects both the diagnosis and the warranty.
Where do the facts on this page come from?
- Public technical literature (including Application, installation and selection of weighing modules): the symptoms and causes of the eight faults, the restraint and piping requirements, and the drift diagnosis.
- Company installation instructions: free movement of the support bolt and sleeve, no welding current through the load cell, no tension on the cable, and no mixing of capacities in one system.
- What this page adds: the cost-ordered check table, the specific confirmation step for each fault, the prevention list and the "call the manufacturer" boundary. These three parts organise the sources above; they do not change them.
This page is general technical guidance, not a commitment for a specific project. On site, the shipping documentation of the actual unit and the real conditions govern.
Still not solved? What should you tell us so we can diagnose it?
What we need: module model and number of modules, indicator model, how long it has been installed, when it started (from day one / after some time / after rain / after welding), vessel dead weight and product weight, rigid or flexible pipe, and whether the line can be stopped.