Short answer: the photograph shows the bare load cell, not an assembly with mounting plates and a top plate. In this factory's naming, M usually denotes a module, but this model's photographs show the cell alone (alloy-steel bar body with a cable exit), with none of the module's top and bottom hardware.
⚠️ This matters because "module or bare cell" decides the whole mounting arrangement. The company text never uses the word module either — it says only "compact structure, good mounting compatibility". If you need the module form supplied, say so before ordering (extra mounting parts may be a separate item).
⚠️ Sensitivity 1.94, not 2.0
Most load cells on this site are 2.0 or 2±0.2 mV/V; this one is 1.94±0.002. Two practical consequences:
- When replacing: putting a 1.94 cell where a 2.0 was (or the reverse) changes the output by roughly 3% at the same load — the indicator must be recalibrated; it cannot simply be swapped in.
- When paralleling: cells on one scale with different sensitivity share the load in proportion to their sensitivity, not equally.
So this belongs on the checklist for indicator matching, replacement and paralleling. This page does not rule on "can it be swapped directly"; it states the difference plainly.
How is it wired?
Short answer: 6-wire. The text and the drawing agree wire for wire.
| Colour | Function | Terminal name |
|---|---|---|
| Green | + excitation | Excitation(+) |
| Black | − excitation | Excitation(−) |
| White | + signal | Signal(+) |
| Red | − signal | Signal(−) |
| Yellow | + sense | Sense(+) |
| Blue | − sense | Sense(−) |
| pale-yellow | shield | Shield |
⚠️ Here it is green = excitation+, white = signal+, red = signal−. That is not the same as the 4-wire HY-1241 on this site (red = excitation+, green = signal+), so never carry a colour order across models. Earth the shield at one end only; at 9 m this is a long run and the shield earthing matters.
What are the specifications?
Short answer: the company documentation gives 19 parameters, transcribed below.
| Parameter | Unit | Value |
|---|---|---|
| Sensitivity | mV/V | 1.94±0.002 |
| Non-repeatability | %R.C. | ≤0.01 |
| Hysteresis | %R.C. | ≤0.02 |
| Non-linearity | %R.C. | ≤0.02 |
| Creep (30 min) | %R.C. | ≤0.02 |
| Zero output | %R.C. | ≤1 |
| Compensated temperature range | °C | -10 to +40 |
| Operating temperature range | °C | -20 to +60 |
| Temperature effect on sensitivity | %Load/°C | ≤0.002 |
| Temperature effect on zero | %R.C./°C | ≤0.002 |
| Input resistance | Ω | 381±4 |
| Output resistance | Ω | 350±1 |
| Insulation resistance | MΩ | ≥5000 (50 V DC) |
| Recommended excitation | V (DC/AC) | 5 to 15 |
| Maximum excitation | V (DC/AC) | 20 |
| Safe overload | %R.C. | 150 |
| Ultimate overload | %R.C. | 300 |
| Protection | — | IP68 |
| Cable length | m | 9 |
IP68 covers continuous immersion, which is steadier than IP67 where a batching area is washed down and the floor stands wet.
What are the dimensions?
Short answer: no values are copied here. The drawing carries a capacity-grouped dimension table (titled "0743 9-13.6t"), but the type is only a few pixels high — enlarged, the strokes merge and the digits cannot be told apart. And only the 9–13.6 t sheet exists; the 20 t drawing is missing. Both facts are recorded here.
| Item | Situation |
|---|---|
| Dimension sheets | 1 (image 03, titled "0743 9-13.6t") |
| Capacity table present? | yes (about 16 columns, two data rows) |
| Values legible? | no (still merged at 8× enlargement) |
| 20 t drawing available? | no (title stops at 13.6 t) |
⚠️ Why not copy: a dimension drawing exists to be drilled from, and one misread number becomes a scrapped base plate — far worse than "no dimension yet". The field is left empty. Ask us for the drawing of this model and say which step (9 t / 13.6 t / 20 t) — the 20 t drawing has to be supplied separately.
What does it look like?
Short answer: four images are supplied — one product photograph, one dimension drawing (9–13.6 t) and one wiring diagram. Three are used here.
⚠️ There is no 20 t dimension drawing — the supplied sheet stops at 13.6 t. Ask separately if you need it.
Where does it fit, and where does it not?
Short answer: the documented applications are ground-level platforms and batching in chemical, food and pharmaceutical plants, 9 to 20 t.
The equipment and industries named above are the documented applications — where this cell goes — not a claim that this factory builds complete platforms.
- Fits: high-capacity batching; chemical, food and pharmaceutical dosing; sites needing IP68 (continuous immersion).
- Why IP68: batching areas are washed down and floors stand wet; IP67 only covers brief immersion.
- ⚠️ Two things to confirm: ① sensitivity 1.94 — recalibrate on replacement or when paralleling; ② this is the bare cell, no mounting hardware — say so early if you need the module form.
- Not covered: small capacities (single point); a wider temperature range (this one stops at +60 °C); 4-wire indicators.
How can I tell whether the cell has failed?
Short answer: input across green/black 381±4 Ω, output across white/red 350±1 Ω, insulation to body ≥5000 MΩ.
| Step | What to do | What to look for |
|---|---|---|
| 1 | Power off; measure green / black | Around 381±4 Ω; open circuit means a broken internal wire |
| 2 | Measure white / red | Around 350±1 Ω |
| 3 | Measure insulation to body | MΩ range (≥5000); a clear drop means moisture or a damaged seal |
Resistances and sensitivity are different things: normal resistances do not prove the sensitivity is right. Check sensitivity against the laser marking or the shipping documentation.
Where do the facts come from?
- Company product documentation: five features, 6-wire colour code, 3 capacity steps, 19 technical parameters.
- Company images (four): one product photograph, one dimension drawing for 9–13.6 t with the capacity table, one wiring diagram.
- Sensitivity check against other models: the other product pages (most state 2.0 or 2±0.2).
- Public technical literature: how sensitivity differences affect parallel sharing and calibration, IP67 versus IP68.
Dimensions are omitted because the table values are illegible and no 20 t sheet exists.
Still unsure which step fits?
Tell us: load per cell and cell count, platform or vessel mounting, whether you need the module form with mounting hardware, indicator model (6-wire?), washdown conditions, and which capacity step you need.