What is it, and how does it relate to HY-SBD?
Short answer: the photograph shows the same double shear beam structure as HY-SBD (beam body, black mounting base, two load buttons), and the capacity range is identical (15 to 50 t). But the electrical parameters differ — this is not the same data sheet under a new name.
| Item | HY-QS (this page) | HY-SBD |
|---|---|---|
| Capacity | 15 to 50 t, 5 steps | 15 to 50 t, 5 steps (same) |
| Input / output resistance | 750±10 / 703±2 Ω | 772±4 / 700±1 Ω |
| Operating range | -30 to +70 °C | -40 to +65 °C |
| Excitation (rec. / max) | 10-12 V / 15 V | 5-15 V / 20 V |
| Overload (safe / ultimate) | split by capacity (see below) | 150% / 300% |
| Protection | IP67 (5 t) / IP68 (10-50 t) | IP67 |
| Cable | 13 / 16 m | 13 m |
⚠️ Two points: ① the excitation window is narrow (10-12 V recommended, 15 V max) compared with HY-SBD's 5-15 / 20 V — an indicator set above 15 V will exceed it; ② the resistances differ (750/703 vs 772/700), so fault checks must not borrow the other model's reference values.
What are the specifications? ⚠️ Three rows are split by capacity
Short answer: the company documentation gives 19 parameters, and three of them carry two values in one cell — safe overload, ultimate overload and protection. Both values are transcribed, with the capacity group the documentation names.
| Parameter | Unit | Value |
|---|---|---|
| Sensitivity | mV/V | 2±0.002 |
| Non-repeatability | %R.C. | ≤0.01 |
| Hysteresis | %R.C. | ≤0.02 |
| Non-linearity | %R.C. | ≤0.02 |
| Creep (30 min) | %R.C. | ≤0.02 |
| Zero output | %R.C. | ≤1 |
| Compensated temperature range | °C | -10 to +40 |
| Operating temperature range | °C | -30 to +70 |
| Temperature effect on sensitivity | %Load/°C | ≤0.002 |
| Temperature effect on zero | %R.C./°C | ≤0.002 |
| Input resistance | Ω | 750±10 |
| Output resistance | Ω | 703±2 |
| Insulation resistance | MΩ | ≥5000 (50 V DC) |
| Recommended excitation | V (DC/AC) | 10-12 |
| Maximum excitation | V (DC/AC) | 15 |
| Safe overload | %R.C. | 150; 120 (40-50 t) |
| Ultimate overload | %R.C. | 300; 200 (40-50 t) |
| Protection | — | IP67 (5 t); IP68 (10-50 t) |
| Cable length | m | 13 / 16 |
⚠️ Two things worth flagging: ① overload capacity drops at the 40-50 t steps (150%→120%, 300%→200%) — the high steps must not be sized with the margin of the low steps; ② the protection cell says "IP67 (5 t)" but the capacity list has no 5 t step (only 15-50 t). That is an inconsistency in the documentation; it is transcribed as written, and this page does not decide whether a 5 t version still exists. Ask us if you need one.
How is it wired?
Short answer: 4-wire. The text and the drawing agree wire for wire; the drawing adds pale-yellow = Shield.
| Colour | Terminal | Function |
|---|---|---|
| Red | +input | + excitation |
| Black | −input | − excitation |
| Green | +output | + signal |
| White | −output | − signal |
| pale-yellow | Shield | Shield (drawing only) |
⚠️ Set the excitation for this model: 10-12 V recommended, 15 V max. Carrying over the 15 V-plus habit from other high-capacity models would exceed this one's limit. Earth the shield at one end only; at 13 m or 16 m this is a long 4-wire run, so assess the drop.
What are the dimensions?
Short answer: no dimensions are published — the documentation does not contain a dimension drawing for this model. Of the six images supplied, four are product photographs and two are wiring diagrams (duplicates); none carries dimension callouts.
| Type | Count | Dimensions? |
|---|---|---|
| Product photographs | 4 | no |
| Wiring diagrams | 2 (duplicates) | no |
| Dimension drawing | 0 | — |
⚠️ Why not estimate: dimensions are used to drill base plates and mounting holes, and an estimated number moves the risk onto the customer. The field is left empty — ask us for the drawing of this model (number in the footer).
What does it look like?
Short answer: this page uses one product photograph and one wiring diagram. Three further photographs from the same angle and one duplicate diagram are omitted.
⚠️ No dimension drawing exists for this model — all six supplied images are photographs or wiring diagrams.
Where does it fit, and where does it not?
Short answer: the documented applications are truck weighing and rail weighing, 15 to 50 t.
The platform types named above are the equipment the cell goes onto, as written in the documentation — not a claim that this factory builds complete platforms.
- Fits: large vehicle-weighing platforms and rail weighing; parallel banks where each cell carries 15 to 50 t.
- Why this model: ① operating range -30 to +70 °C, wider than the sibling (HY-SBD is -40 to +65 °C); ② the 10-50 t steps are documented IP68 (continuous immersion), steadier than IP67 outdoors.
- ⚠️ Three limits: ① excitation max 15 V (10-12 recommended); ② overload margin drops at 40-50 t (120% / 200%); ③ no dimension drawing — work from the unit or ask for the drawing.
- Not covered: a lower temperature floor (HY-SBD reaches -40 °C); small capacities (single point or column); long runs needing sense wires (this model is 4-wire).
How can I tell whether the cell has failed?
Short answer: input across red/black 750±10 Ω, output across green/white 703±2 Ω, insulation to body ≥5000 MΩ.
| Step | What to do | What to look for |
|---|---|---|
| 1 | Power off; measure red / black | Around 750±10 Ω; open circuit means a broken internal wire |
| 2 | Measure green / white | Around 703±2 Ω |
| 3 | Measure insulation to body | MΩ range (≥5000); a clear drop means moisture or a damaged seal |
⚠️ Do not borrow HY-SBD's references (772/700) — the resistances differ and the wrong reference will fail a good cell. On a parallel bank, disconnect the other cells before measuring one.
Where do the facts come from?
- Company product documentation: five features, 4-wire colour code, 5 capacity steps, 19 technical parameters (including three capacity-split rows).
- Company images (six): four product photographs, two wiring diagrams (duplicates); no dimension drawing.
- Check against HY-SBD: both pages (same capacity and structure; different electrical parameters).
- Public technical literature: double shear beam load paths, excitation drop over long cables, IP67 versus IP68.
The protection row says "IP67 (5 t)" while the capacity list has no 5 t step; both are transcribed unchanged.
Still unsure which step fits?
Tell us: platform size and cell count, total capacity per cell, foundation type, distance to the indicator, what excitation voltage your indicator supplies (this model wants 10-12 V, max 15 V), and whether the site stands wet.