HY-0745A-M Load Cell: 0.22 t to 4.4 t, 6-Wire, IP68, 4.5 m Cable

HY-0745A-M is a medium-to-small capacity load cell, 0.22 t to 4.4 t in 5 steps, for ground-level platforms and batching in chemical, food and pharmaceutical plants. Compact body, good mounting compatibility, 6-wire with a sense pair, IP68, 4.5 m cable. The drawing carries a clearly legible capacity table, and this page reproduces it in full.

Not sure about the step:

+86 183 5123 8137

Short answer: HY-0745A-M covers 0.22, 0.55, 1.1, 2.2 and 4.4 t; compact body, good mounting compatibility, IP68; wiring is 6-wire (green = +excitation, black = −excitation, white = +signal, red = −signal, yellow = +sense, blue = −sense); sensitivity 1.94±0.002 mV/V, input 381±4 Ω, output 350±1 Ω, operating range -20 to +60 °C, safe overload 150%, ultimate 300%, cable 4.5 m.

How does it split from HY-0743-M?

Short answer: the features, the 6-wire colour order and all 19 parameters are word-for-word identical. Only two things differ: the capacity ladder and the cable length. This one covers the smaller range (0.22–4.4 t); HY-0743-M covers 9–20 t.

The two siblings (both pages of documentation)
Item HY-0745A-M (this page) HY-0743-M
Capacity 0.22 to 4.4 t, 5 steps 9 to 20 t, 3 steps
Sensitivity 1.94±0.002 1.94±0.002 (same)
Other 18 parameters identical identical
Features 5 lines, word-for-word same 5 lines, word-for-word same
Wiring 6-wire, same colour order 6-wire, same colour order
Cable 4.5 m 9 m

⚠️ Selection comes down to one number: the load per cell. 0.22–4.4 t takes this model; 9–20 t takes HY-0743-M. Both are 1.94, so they are electrically interchangeable (still re-calibrate), but their capacity ranges are far apart and they cannot substitute for one another.

How is it wired?

Short answer: 6-wire. The text, the label on the unit and the wiring diagram all agree.

6-wire colour code (text / unit label / drawing agree)
Colour Function Terminal name
Green+ excitationExcitation(+)
Black− excitationExcitation(−)
White+ signalSignal(+)
Red− signalSignal(−)
Yellow+ senseSense(+)
Blue− senseSense(−)
long yellowshieldShield

⚠️ The shield has now appeared in three different wordings across this factory's documents: this unit's label says "long yellow", the HY-SBT label says "thick yellow", and the drawings say pale-yellow. All mean the same thing; the actual wire governs. Earth the shield at one end only.

What are the specifications?

Short answer: the company documentation gives 19 parameters, transcribed below.

HY-0745A-M specifications (company documentation)
Parameter Unit Value
SensitivitymV/V1.94±0.002
Non-repeatability%R.C.≤0.01
Hysteresis%R.C.≤0.02
Non-linearity%R.C.≤0.02
Creep (30 min)%R.C.≤0.02
Zero output%R.C.≤1
Compensated temperature range°C-10 to +40
Operating temperature range°C-20 to +60
Temperature effect on sensitivity%Load/°C≤0.002
Temperature effect on zero%R.C./°C≤0.002
Input resistanceΩ381±4
Output resistanceΩ350±1
Insulation resistanceMΩ≥5000 (50 V DC)
Recommended excitationV (DC/AC)5 to 15
Maximum excitationV (DC/AC)20
Safe overload%R.C.150
Ultimate overload%R.C.300
Protection—IP68
Cable lengthm4.5

IP68 covers continuous immersion, steadier than IP67 where a batching area is washed down.

What are the dimensions? ⭐ This drawing table is legible

Short answer: the drawing carries a clearly legible capacity table (titled "0745A"), grouped as 0.22–1.1 t / 2.2 t / 4.4 t with 14 columns. It is reproduced in full below. Units are not marked on the drawing and none are added here.

Capacity-grouped dimensions (from drawing "0745A", units not marked)
Capacity / t A B C D E F G H K R M P Q S
0.22–1.1 153.915.476.225.430.235.757.717.34.512.722.215.93.17.9
2.2 158.615.476.225.430.636.957.822.98.212.722.215.93.112.7
4.4 171.521.495.235.142.942.973.828.310.819.0534.922.2017.8

⚠️ Read the three rows carefully: 0.22 / 0.55 / 1.1 t share one set; 2.2 t is its own row, and 4.4 t another, with A (overall length) growing 153.9 → 158.6 → 171.5. Use the row for your capacity — do not fit a 4.4 t unit with 0.22 t hole spacing. The drawing also marks the relief cut-out and the sealant area.

What does it look like?

Short answer: five images are supplied — product photographs, one dimension drawing with the capacity table, and one wiring diagram. This page uses one photograph plus both drawings.

HY-0745A-M load cell: natural alloy-steel bar body with mounting holes, a wiring label on the side (green +excitation, black -excitation, white +signal, red -signal, yellow +sense, blue -sense, long yellow shield); factory watermark
The label on the body is a wiring instruction, not a nameplate, and matches the colour table above. Company documentation.
HY-0745A-M dimension drawing (title 0745A): side and top callouts, plus a capacity-grouped 14-column table with three rows (0.22-1.1 t, 2.2 t, 4.4 t)
Dimension drawing (capacity table reproduced in full above). Company documentation.
HY-0745A-M wiring diagram: Green excitation+, Yellow sense+, White signal+, Black excitation-, Blue sense-, Red signal-, pale-yellow shield
Wiring diagram (6-wire plus shield). Company documentation.

⚠️ This is one of the few legible capacity tables in the factory's material (most are too small to read), so all 14 columns are reproduced here for direct use when drilling.

Where does it fit, and where does it not?

Short answer: the documented applications are ground-level platforms and batching in chemical, food and pharmaceutical plants, 0.22 to 4.4 t.

The equipment and industries named above are the documented applications — where this cell goes — not a claim that this factory builds complete platforms.

How can I tell whether the cell has failed?

Short answer: input across green/black 381±4 Ω, output across white/red 350±1 Ω, insulation to body ≥5000 MΩ.

Three checks that can be done on site
Step What to do What to look for
1 Power off; measure green / black Around 381±4 Ω; open circuit means a broken internal wire
2 Measure white / red Around 350±1 Ω
3 Measure insulation to body MΩ range (≥5000); a clear drop means moisture or a damaged seal

Resistances and sensitivity are different things: normal resistances do not prove the sensitivity is right. Check sensitivity against the marking or the shipping documentation.

Where do the facts come from?

Values are transcribed from company documentation and are not a commitment for a specific project.

Still unsure which step fits?

Tell us: load per cell and cell count, platform or vessel mounting, indicator model (6-wire?), washdown conditions, and mounting space and hole pattern.

Call +86 183 5123 8137