HY-1241 Load Cell: Single Point, 30 kg to 250 kg, 4-Wire

HY-1241 is a single-point load cell for bench scales, packing scales and process weighing. Unlike the column-type HY-MTB range, a single-point cell has its corner errors compensated at the factory — a load moved around the platform does not need corner-by-corner correction on site. Laser-welded seal; 2 m cable.

Not sure between single-point and column-type:

+86 183 5123 8137

Short answer: HY-1241 covers 30, 50, 100, 200 and 250 kg in 5 steps; it is a single-point cell, laser-welded, with corner errors factory-compensated; wiring is 4-wire (red = +excitation, black = −excitation, green = +signal, white = −signal); sensitivity 2±0.2 mV/V, input 400±10 Ω, output 350±4 Ω, safe overload 150%, ultimate 300%, cable 2 m.

⚠️ First, a conflict in the documentation: is the protection rating IP65 or IP67?

Short answer: the company documentation states it twice, and the two statements disagree. Both are transcribed here; neither is chosen for the customer.

Two statements in the same company document
Where What it says
Feature list "Laser-welded seal, protection rating IP65"
Parameter table "Protection: IP67"

⚠️ This is not a difference between two models — it is a direct conflict inside one page. The practical difference sits in short-term immersion: IP67 tolerates it, IP65 does not. This page does not decide which one is right: where there is washdown, standing water or possible immersion, go by the shipping documentation and the nameplate on the unit, or call to confirm.

What is it, and how does it differ from the column type?

Short answer: a single-point cell — a machined body with cut-outs that deflect under load — intended for platform loading, such as bench scales and packing scales.

Single-point vs column type (two ranges from this factory)
Item HY-1241 (single point) HY-MTB / HY-MTB-M (column)
Load path One point, body deflection Column in compression, with spherical joint and support bolts
Corner errors Compensated at the factory Handled by mounting and corner calibration
Typical equipment Bench scales, packing scales, process weighing Packing scales, belt scales, batching
Capacity 30 to 250 kg (5 steps) 5 to 500 kg (10 steps)

How is it wired?

Short answer: 4 wires (red / black / green / white). The product page text lists only those four; the wiring diagram on the same page additionally marks pale-yellow = Shield.

4-wire colour code (company documentation)
Colour Terminal Function
Red+input+ excitation
Black−input− excitation
Green+output+ signal
White−output− signal
pale-yellowShieldShield (drawing only; not listed in the text)

⚠️ One documentation difference, transcribed without merging: the text lists four wires and does not mention a shield; the diagram marks one. Earth the shield at one end only.

What are the specifications?

Short answer: the company documentation gives 19 parameters, transcribed below. Note that the values differ from the column-type HY-MTB range.

HY-1241 specifications (company documentation)
Parameter Unit Value
SensitivitymV/V2±0.2
Non-repeatability%R.C.≤0.01
Hysteresis%R.C.≤0.02
Non-linearity%R.C.≤0.02
Creep (30 min)%R.C.≤0.02
Zero output%R.C.≤10
Compensated temperature range°C-10 to +40
Operating temperature range°C-20 to +65
Temperature effect on sensitivity%Load/°C≤0.002
Temperature effect on zero%R.C./°C≤0.002
Input resistanceΩ400±10
Output resistanceΩ350±4
Insulation resistanceMΩ≥5000 (50 V DC)
Recommended excitationV (DC/AC)5 to 15
Maximum excitationV (DC/AC)20
Safe overload%R.C.150
Ultimate overload%R.C.300
Protection—IP67 (⚠️ conflicts with IP65 in the feature list)
Cable lengthm2

How do these numbers differ from the column-type HY-MTB?

Differences between the two ranges (transcribed)
Parameter HY-1241 (single point) HY-MTB (column)
Sensitivity2±0.22±0.002
Zero output≤10≤1
Operating range-20 to +65-40 to +65
Input resistance400±10387±4
Output resistance350±4350±1
Cable2 m3 m

These are not the same parameter set under a different name: the sensitivity tolerance is 100× wider (±0.2 vs ±0.002), zero output is 10× larger, and the operating floor is 20 °C higher. This page lists the facts and does not decide "which is more accurate".

What are the dimensions?

Short answer: one dimensioned drawing. Units are not marked and none are added here.

Dimensions transcribed from the drawing
Direction Values shown
Overall length174
Height / width65 (side view height), 60 (top view width)
Hole spacings122, 19, 19, 7; holes 2-Φ8.5

The drawing also carries "15 (4 holes)" and "30 (2 holes)" callouts and the single-point over-load relief cut-outs.

What does it look like?

Short answer: the documentation contains four photographs and two drawings. This page shows the two drawings only: the nameplates on the photographs are legible, but they carry marks beyond the published product parameters, and the images carry a factory watermark, so they are not published here.

HY-1241 wiring diagram: red excitation+, black excitation-, green signal+, white signal-, plus pale-yellow shield
Wiring diagram (4-wire plus shield). Company drawing.
HY-1241 dimension drawing: overall 174, height 65, width 60, spacings 122/19/19/7, holes 2-Φ8.5
Outline drawing (single point, with relief cut-outs). Company drawing.

No usable product photograph is published for this model. Where an image and the written documentation disagree, the written documentation governs.

Where does it fit, and where does it not?

Short answer: the documented applications are bench scales, packing scales and process weighing. Anything involving washdown or standing water should have the protection-rating conflict above settled first.

How can I tell whether the cell has failed?

Short answer: three measurements — input across red/black 400±10 Ω, output across green/white 350±4 Ω, insulation to the body ≥5000 MΩ.

Three checks that can be done on site
Step What to do What to look for
1 Power off; measure red / black Around 400±10 Ω; open circuit means a broken internal wire
2 Measure green / white Around 350±4 Ω
3 Measure insulation to body MΩ range (≥5000); a clear drop means moisture or a damaged seal

If the values are normal but the reading is unstable, check installation, piping, junction box and supply first — see Selection & Resources.