HY-MTB-M Load Cell: 5 kg to 500 kg, Stainless Steel, IP68

HY-MTB-M is a stainless-steel column-type load cell with a low mounting height, for packing scales, belt scales and batching control in chemical, medical and food plants. The body is hermetically welded and rated IP68; the cable leaves from one end.

Not sure which capacity:

+86 183 5123 8137

Short answer: HY-MTB-M covers 5, 10, 20, 30, 50, 75, 100, 200, 300 and 500 kg in 10 steps; the body is stainless steel, hermetically welded, IP68; sensitivity 2±0.002 mV/V, input resistance 387±4 Ω, output resistance 350±1 Ω, insulation resistance 5000 MΩ, safe overload 150%, ultimate overload 300%, cable 3 m. The company documentation also states that a high-temperature version is available, with a maximum operating temperature of 150 °C.

What is it, and where is it used?

Short answer: a column-type load cell that converts load into a millivolt signal for an indicator or transmitter. Its documented features are low mounting height and a stainless-steel hermetically welded body, which suit tight spaces that also need protection from moisture.

The company documentation names packing scales and belt scales, plus batching control in chemical, medical and food plants. What these have in common: small loads, high cycle counts, and washdown or dust on site.

Load cell vs weighing module: a load cell is the element; a module adds the load-bearing structure and mounting connections, and takes care of levelling and restraint. This page covers the former; see fixed weighing module for the latter.

What are the specifications?

Short answer: the company documentation gives 19 parameters, transcribed below with their units and symbols, without conversion or rounding.

HY-MTB-M specifications (company documentation)
Parameter Unit Value
SensitivitymV/V2±0.002
Non-repeatability%R.C.≤0.01
Hysteresis%R.C.≤0.02
Non-linearity%R.C.≤0.02
Creep (30 min)%R.C.≤0.02
Zero output%R.C.≤1
Compensated temperature range°C-10 to +40
Operating temperature range°C-40 to +65
Temperature effect on sensitivity%Load/°C≤0.002
Temperature effect on zero%R.C./°C≤0.002
Input resistanceΩ387±4
Output resistanceΩ350±1
Insulation resistanceMΩ5000 (50 V DC)
Recommended excitationV (DC/AC)5 to 15
Maximum excitationV (DC/AC)20
Safe overload%R.C.150
Ultimate overload%R.C.300
Protection—IP68
Cable lengthm3

What do the key parameters mean on site?

What each key parameter determines
Parameter What it determines
Safe 150% / ultimate 300% Safe overload is what the cell can carry continuously; ultimate is a one-off limit. Neither is for routine overload — size so that each cell runs at about 70% of its rating (see sizing).
Input 387±4 Ω / output 350±1 Ω The reference values for a field check: measure these two terminal pairs with a multimeter; a large deviation suggests an open circuit or moisture ingress.
Insulation 5000 MΩ (50 V DC) The key indicator for moisture. Once insulation drops, readings drift or jump.
Excitation 5–15 V, max 20 V The indicator supply must sit inside this window: too high heats the cell, too low leaves the signal weak.
Compensated -10 to +40 °C, operating -40 to +65 °C Compensation is the range in which the cell was trimmed; operation is what it tolerates. Outside the compensated range, temperature effects grow — recalibrate for the duty if needed.
Cable 3 m Supplied length, not spare slack. Longer runs need an extension, with the shield and joints handled properly.

How is it wired?

Short answer: 6-wire, and the product page text matches the wiring diagram wire for wire.

6-wire colour code
Colour Function Terminal name
Green+ excitationExcitation(+)
Black− excitationExcitation(−)
White+ signalSignal(+)
Red− signalSignal(−)
Yellow+ senseSense(+)
Blue− senseSense(−)
pale-yellowshieldShield

The sense pair compensates excitation loss over long cable runs. Earth the shield at one end only; both ends create a ground loop.

What are the dimensions?

Short answer: the company documentation provides two dimensioned drawings, one for MTB5-10kg and one for MTB20-300kg. No drawing is provided for the 500 kg step. Units are not marked on the drawings and none are added here.

Dimensions transcribed from the drawings
Drawing Covers Values shown
Drawing 2 MTB5-10kg (5 kg, 10 kg) Length 123, 82, 45, 26; end view diameter Φ35, height 20; hole callouts; spacings 18 and 10
Drawing 3 MTB20-300kg (20 to 300 kg) Length 123, 82, 45, 26, 42; end view diameter Φ36, height 20; holes 2-Φ8.2 and Φ8.2; spacings 18 and 10
None 500 kg No drawing supplied; not inferred here

Note: the company product page also carries a three-row dimension table whose diameter symbol was corrupted during capture (e.g. "?45", "?8.2"). This page does not use the corrupted characters; it relies on the readable drawing callouts above.

What does it look like?

Short answer: the documentation contains two photographs and three drawings. This page shows the three drawings only: the nameplates on the two photographs, read at high magnification, contain items not published here by this site's image policy.

HY-MTB-M wiring diagram: bridge schematic with Green excitation+, Yellow sense+, White signal+, Black excitation-, Blue sense-, Red signal-, and pale-yellow shield
Wiring diagram: Green / Yellow / White / Black / Blue / Red / pale-yellow (shield). Company drawing.
HY-MTB-M dimension drawing for MTB5-10kg: end and top views with 123, 82, 45, 26, 20, Φ35, 18 and 10
MTB5-10kg dimension drawing (5 kg and 10 kg). Company drawing.
HY-MTB-M dimension drawing for MTB20-300kg: end and top views with 123, 82, 45, 26, 42, 20, Φ36, 2-Φ8.2, 18 and 10
MTB20-300kg dimension drawing (20 to 300 kg). Company drawing.

No usable product photographs are published for this model; the two supplied photographs were not used. Where an image and the written documentation disagree, the written documentation governs.

Where does it fit, and where does it not?

Short answer: the documented applications are packing scales, belt scales and batching control in chemical, medical and food plants, and a high-temperature version up to 150 °C is available. Anything beyond those is not inferred here.

How can I tell whether the cell has failed?

Short answer: measure three values and compare with the table above — input 387±4 Ω, output 350±1 Ω, and insulation in the MΩ range.

Three checks that can be done on site
Step What to do What to look for
1 Power off, measure resistance across excitation (green / black) Around 387±4 Ω; open circuit or large deviation means a broken internal wire
2 Measure across signal (white / red) Around 350±1 Ω
3 Measure insulation between conductors and body MΩ range; a clear drop means moisture or a damaged seal

If the values are normal but the reading is unstable, work through the installation, piping, junction box and supply checks on the Selection & Resources page before replacing the cell.

Where do the facts come from?

Values are transcribed from company documentation. The documentation contains one protection-type wording carried over from earlier material; this page does not reproduce that rating and makes no claim based on it.

Still unsure which step fits?

Tell us: equipment type (packing scale / belt scale / batching scale), load per cell, mounting and space, material and site conditions (dust, washdown, temperature), whether a high-temperature or explosion-proof version is needed, and the distance to the indicator.