What is HY-FW-M and where does it mount?
Short answer: a weighing module is a structural accessory for load cells, made of a top plate, a base plate, the load cell, a load-bearing column and support bolts; it combines the load cell, the load-transfer parts and the mounting plate in one unit. (Quoted from the public technical paper "Selection and Installation of Weighing Modules".)
Modules usually mount on the lower part of the scale body (feet, platform), and can also mount on the upper part of the body (the waist of a reactor, tank or hopper). All modules of one weighing system must sit in the same horizontal plane. (Public technical literature.)
The company documentation describes HY-FW-M as: unique structure, easy to mount on many kinds of tank; support bolts prevent the equipment from tipping; simple, quick installation; easy maintenance, less downtime. The matching installation step is: mount the module under the tank foot or bracket, then fix the module to the foot with the support bolts. (Company product documentation.)
What are the three module structures, and which one is this?
Per public industry literature, weighing modules are classed as fixed, semi-floating and floating according to the load-bearing column and the top plate. In one system you use one fixed module and one semi-floating module, and fill the rest with floating modules; sets are usually three modules (three-point support) or four modules (four-point support). (Public technical literature.)
The company documentation does not state which of the three structures HY-FW-M is, and this page does not classify it. The division of work and the criteria are on the fixed weighing module page; for a structure-based selection use that page and the shipping documentation. (Basis: the structure class of this model is not recorded in the company documentation.)
Where does the module sit in a weighing system?
A weighing control system is usually made of the indicator, the junction box, the weighing modules and the cable; the modules carry the load, the junction box sums the signals of several modules, and the indicator displays and controls. (Company installation instructions.)
One system usually uses 3 to 6 sets (12 sets maximum) of modules as a group; the foundation must meet the stiffness requirement and sit in the same plane (or be adjustable to it), so that every module carries its share evenly. (Company installation instructions.)
How do HY-FW, HY-FW-M and HY-FWC-M differ?
Short answer: the three are all weighing modules for tank batching, and the documented main features are almost the same. The real differences sit in two places — wiring and material / protection. HY-FW is 4-wire + shield; HY-FW-M (this page) is 6-wire + shield with the same alloy steel, nickel-plated body; HY-FWC-M is also 6-wire + shield but uses a stainless-steel hermetically sealed load cell, and IP68 is documented for it. Capacity differs too: HY-FW and HY-FW-M both cover 0.5 t to 20 t (9 steps), HY-FWC-M covers 0.22 t to 20 t (7 steps).
| Model | Wiring | Material and sealing | Protection | Capacity |
|---|---|---|---|---|
| HY-FW | 4-wire + shield | Alloy steel, nickel-plated | Not stated in the company documentation | 0.5 t to 20 t (9 steps) |
| HY-FW-M (this page) | 6-wire + shield | Alloy steel, nickel-plated | Not stated in the company documentation | 0.5 t to 20 t (9 steps) |
| HY-FWC-M | 6-wire + shield | Stainless-steel hermetically sealed load cell | IP68 | 0.22 t to 20 t (7 steps) |
All three rows come from the company product documentation; full specifications for HY-FW and HY-FWC-M are on their own model pages.
Three steps to match your unit
- You have a 4-wire module with a capacity between 0.5 t and 20 t: that is HY-FW (4-wire + shield). (Company product documentation.)
- You need 6 wires, a capacity between 0.5 t and 20 t and an alloy steel, nickel-plated body: that is HY-FW-M, this page. Against HY-FW the wiring is the only change — two wires more, +sense / −sense. (Company product documentation.)
- You must select by material or protection: the company documentation states no protection rating for HY-FW-M, and this page neither infers nor publishes one; when a protection-based decision is needed, use the HY-FWC-M row. (Basis: the protection rating of this model is not recorded in the company documentation.)
Why must the colour code not be copied across models?
The six wires of HY-FW-M are green = +excitation, black = −excitation, white = +signal, red = −signal, yellow = +sense, blue = −sense, and the specification drawing marks pale yellow = shield; HY-FW in the same family is 4-wire and uses a different set. Colour order is not interchangeable across models or batches, so before wiring work use the shipping documentation and the wiring diagram on the actual unit, not colours copied from another model. (Company product documentation and company specification drawing.)
What are the HY-FW-M specifications, and where does each value come from?
Short answer: the table below lists the HY-FW-M values one by one with their source, and the right-hand column turns each value into what it decides at selection and installation. Values are transcribed as they are: no conversion, no merging, no capacity step that the documentation does not list.
| Parameter | Value (with source) | What it determines |
|---|---|---|
| Model | HY-FW-M Company product documentation | Ordering reference. The family also has HY-FW (4-wire + shield) and HY-FWC-M (stainless sealed, IP68); see the table above. Always quote the full suffix. |
| Category | Weighing module Company product documentation | It is a structural part plus a load cell; capacity, material and mounting dimensions all matter together, not the load cell alone. |
| Capacity steps | 0.5, 1, 2, 3, 5, 8, 10, 15, 20 t Company product documentation (9 steps, transcribed verbatim) | The maximum load one module carries. System capacity = rated capacity × number of modules, so calculate from tank dead weight and product weight first, then take the next step up. |
| Material and finish | Alloy steel, nickel-plated Company product documentation | Decides the material and the surface state of the housing and the load-bearing parts; this model is an alloy steel body with a nickel-plated surface. Where the site (damp, washdown, corrosive media, temperature) asks for more, check it against the duty and the shipping documentation — the company documentation gives no rating here, so none is inferred. |
| Overall accuracy | The company documentation states only "a quality load cell gives high weighing accuracy"; no figure is given Company product documentation | The documentation is qualitative only, so no figure is published here and nothing is converted. Where a quantified value is needed, check it against the actual duty and the indicator in use. |
| Structural features | Unique structure, easy to mount on many kinds of tank; support bolts prevent the equipment from tipping; simple, quick installation; easy maintenance, less downtime Company product documentation | "Mounts on many kinds of tank" decides how well it adapts to the foot design; "support bolts prevent tipping" decides that room for fixing and adjusting the support bolts must be left; "easy maintenance" decides the service space under the foot. In practice, measure the existing bolt pattern and the available height first, then choose the capacity step against the two dimension drawings on this page. |
| Wiring system | 6-wire + shield (+excitation / −excitation / +signal / −signal / +sense / −sense, plus one shield wire) Company product documentation and specification drawing | Decides how many wires and terminals the run to the junction box needs. 6-wire adds +sense / −sense over 4-wire: they bring back the excitation voltage lost along the cable, so on long runs, or where the module sits far from the indicator, connect this pair to the sense terminals of the indicator or junction box. The shield suppresses interference from drives and power cables; field practice is to earth the shield at one end and keep signal cables away from power cables. (Practice from public technical literature.) |
| Colour code | Green = +excitation; black = −excitation; white = +signal; red = −signal; yellow = +sense; blue = −sense (6 wires). The wiring diagram on the specification drawing also marks pale yellow = shield Company product documentation and specification drawing | Field wiring reference. Colour order is not interchangeable across models or batches, so use the shipping documentation and the wiring diagram on the actual unit; do not copy colours from another model. |
| Wiring diagram | 6-wire + shield: green = excitation+; yellow = sense+; white = signal+; black = excitation−; blue = sense−; red = signal−; pale yellow = shield Company specification drawing (drawing 3 below) | The colour order on the drawing matches the product-documentation order in the row above wire by wire; check the terminal numbers against the drawing before powering up, and where the two disagree, the shipping documentation and the unit itself govern. |
| Outline dimensions | Two dimension drawings (side view and end view) with 15 designators: A, A1, 4-ØG, H, T, L1, L2, L, 4-ØP (side view, where 4-ØG and 4-ØP are hole positions) plus B, B1, H1, H2, C, C1 (end view). This model has no per-step dimension table, so no dimension values are published here; the drawings state no unit and are transcribed as they are Company specification drawing (drawings 4 and 5 below) | Whether a capacity step also changes the outline and the mounting holes has to be checked on the actual unit. Fix the capacity step before modifying feet or casting a foundation, then check the hole pattern, the overall height and the available space. For the value behind each designator, use the specification drawing and the shipping documentation. |
| Documented applications | Batching control on tanks Company product documentation | The documentation gives a direction for the duty, not a promise that any tank will do. For a given vessel, check the vessel weight, the product weight, the number of feet and the site conditions. |
Eleven parameters are listed, matching the eleven fields recorded in the company documentation. Values and wording are transcribed as they are, including the fact that no figure is given for some of them; every explanation this page adds carries its own source and comes from public technical literature rather than from a company commitment. The company documentation gives no protection rating and no long-term stability statement for this model, so neither is published or inferred here.
Which capacity step fits which duty?
Short answer: check the rated capacity of one module against "dead load + live load ≤ rated capacity × number of modules × 70%", then take the next step up; the number of modules follows the number of real support points. (Criteria from public technical literature.)
dead load + live load ≤ rated capacity × number of modules × 70%
The formula is repeated as text so it can be copied into a design note or printed for a design office. Worked examples are on the Selection & Resources page.
Case by case (follow it directly)
| Your case | What to do | Source |
|---|---|---|
| The tank already has support feet | One module per support foot: 3 feet, 3 modules; 4 feet, 4 modules | Public technical literature |
| The calculated capacity per module falls between two steps | Take the next step up, never down | Public technical literature |
| How many modules to use, and which structure | One fixed module, one semi-floating module, the rest floating; three-module and four-module sets are the usual arrangements | Public technical literature |
| Product is fed into the vessel horizontally | Horizontal feed creates horizontal impact force, which calls for a dynamic-load module; vertical feed calls for a static-load module | Public technical literature |
| The module sits far from the indicator | Use the +sense / −sense pair of the 6-wire system to compensate the excitation drop along the cable, wired to the terminals given in the table above | Principle from public technical literature |
| Vessel dead weight and product are both large, fewer modules wanted | Work out the rated capacity per module at 70% first, then set the number of modules from the support points — module count follows the support layout and cannot be replaced by a bigger capacity step | Criteria from public technical literature |
Where does it fit, and where does it not?
Short answer: what the company documentation and the sourced material on this page support is the tank batching control duty, plus the cases the public installation literature points to — for instance horizontal feed creating impact force, which calls for a dynamic-load module. Items the company documentation does not give (protection rating, long-term stability, structure class, material grade, service temperature) are neither inferred here nor turned into a conclusion for the customer.
Fits (documented)
- Batching control on tanks: the documented purpose line is "batching control on tanks". (Company product documentation.)
- Converting an existing vessel, hopper or reactor into a weighing system: the module mounts easily on vessels of many shapes, which suits retrofits of existing equipment. (Public technical literature.)
- Tank support points that need support bolts against tipping: the basis is the documented line "support bolts prevent the equipment from tipping". (Company product documentation.)
- Where product is fed horizontally into the vessel and creates horizontal impact force: feeding vertically by conveyor or pump into a hopper, tank or reactor calls for a static-load module, while horizontal feed onto a platform or into a vessel creates horizontal impact force and calls for a dynamic-load module. (Public technical literature.)
Limits and non-fits (read these too)
- Protection rating, long-term stability, structure class, material grade and service temperature are not given by the company documentation: none of them can be inferred here and no value is offered; where a fixed value is needed, check it separately against the site and the process. (Basis: not recorded in the company documentation.)
- Never mix capacities within one weighing system. (Company installation instructions.)
- No welding current may pass through the load cell: current through the module during welding damages the cell. (Company installation instructions.)
- The vessel must not be mechanically tied to any part the modules do not carry, otherwise the weighing reading is wrong. (Public technical literature.)
- Pipes left as they are make the system impossible to adjust: run all pipes connected to the tank horizontally so they cannot hang the tank, put the first pipe support as far from the vessel as possible, and use flexible connections. (Public technical literature.)
What do the photos and drawings show?
Short answer: one product photo is available for this model, plus five company specification drawings: mounting arrangements, a layout drawing (including the three structure types), the wiring diagram, and two outline-dimension drawings (side view and end view). This model has no per-step dimension table, so dimensions are listed as designators only, with no values. All images are local files; the page loads nothing from outside.
One product photo and five specification drawings, all from the company product documentation and used unaltered. Where an image and the written documentation disagree, the written documentation governs.
What matters during installation and wiring?
Short answer: three things decide whether the module works — the stiffness and level of the base, the verticality of the mounting, and whether outside forces (pipes, welding) have been taken off the load path. Each requirement below carries a value or an action that can be checked, with its source.
| Item | Requirement | Source |
|---|---|---|
| Verticality | The vertical axis of every module must match the perpendicular to the mounting plate; deviation must not exceed 1.5°, and shims correct it | Company installation instructions |
| Frame and base plate stiffness | The frame and base plate carrying the module must be stiff enough; stiffness must reach 1/1000 | Company installation instructions |
| Bracket skew | Skew or twist of the upper bracket and the base bracket against horizontal must not exceed 0.5° | Public technical literature |
| Foundation | The foundation (or floor) carrying the tank must have enough bearing capacity and stiffness; foundation deflection must stay below 3 mm; a concrete foundation is the better choice | Public technical literature |
| Levelling | After mounting, rock the top plate by hand; if it moves, the module is not level — shim with steel plates until it no longer moves | Company installation instructions |
| Bolting | Tighten all bolts diagonally, pre-tighten first, then torque to specification | Company installation instructions |
| Welding | No current may pass through the module during welding; no welding or other current may pass through the load cell | Company installation instructions |
| Cable | Do not pull the cable tight, or tension affects electrical performance; protect the cable from cuts | Company installation instructions |
| Handling | Never lift the module by its cable; avoid knocks and do not let it drop from height | Company installation instructions |
| Fixing | Once the module is level it can be fixed with bolts or by welding (no current through the module while welding) | Company installation instructions |
| Piping | Run all pipes connected to the tank horizontally so they cannot hang the tank; use flexible connections and keep the first support as far from the vessel as possible | Public technical literature |
| Earthing | Each module needs its own independent earth | Public technical literature |
| Wiring and pre-power checks | Wire the load cell cable to the junction box as the wiring diagram requires; re-check every connection and check that the supply voltage is correct | Company installation instructions |
| Calibration conditions | Provide test-weight positions and jacking points for installation and service at the design stage; calibrate after installation | Public technical literature |
The table holds the general requirements of the public installation literature for this class of module; on site, the shipping documentation and the actual duty govern. Indicator-side steps and fault checks are on the FAQ page.
Where do the facts come from?
- Company product documentation (HY-FW-M parameters and main features): model, category, capacity steps, material and finish, the wording of the overall accuracy line, structural features, 6-wire wiring and colour order, purpose line.
- Company specification drawing: mounting arrangements, layout drawing (with the three structure types), the 6-wire with shield wiring diagram, and the side-view and end-view dimension drawings.
- Company installation instructions: make-up of a weighing system and its module count, foundation and stiffness requirements, installation tolerances, the prohibitions on mixing and on welding current, cable handling and protection, wiring and pre-power checks.
- Public technical literature ("Selection and Installation of Weighing Modules" and similar, marked at each sentence that uses it): module make-up and the three structure classes, module count and support arrangements, bracket skew and foundation requirements, piping and calibration requirements, static versus dynamic load criteria, and the 70% sizing rule.
Values in the table are transcribed from company documentation and are not a commitment for a specific project; final specifications are confirmed against the duty. Dates and wording on this page follow the header and footer.