HY-HLJ-M Load Cell: 0.5 t to 3 t, 4-Wire, Self-Stabilising, IP68

HY-HLJ-M is a compact, self-stabilising load cell, 0.5 t to 3 t in 4 steps, for ground-level platforms and batching in chemical, food and pharmaceutical plants. Alloy steel nickel-plated, laser-welded, IP68, 4-wire, 5 m cable. The documentation lists two sensitivity values; both are transcribed below.

Not sure about the step:

+86 183 5123 8137

Short answer: HY-HLJ-M covers 0.5, 1, 2 and 3 t; compact, self-stabilising, alloy steel nickel-plated, laser-welded, IP68; wiring is 4-wire (red = +excitation, black = −excitation, green = +signal, white = −signal); sensitivity is given as two values (2±0.002 / 3.0±0.002); input 381±4 Ω, output 350±1 Ω, operating range -20 to +60 °C, safe overload 150%, ultimate 300%, cable 5 m.

⚠️ First, something the documentation leaves open: why two sensitivity values?

Short answer: the sensitivity cell carries two values — 2±0.002 and 3.0±0.002 — with no statement of which capacity step gets which. Both are transcribed; this page assigns neither.

The sensitivity row as written
Parameter Unit As written
Sensitivity mV/V 2±0.002 / 3.0±0.002

⚠️ Practical consequence: cells with different sensitivity share the load in proportion to their sensitivity when paralleled, not equally. Mixing the two on one scale makes calibration harder. Confirm that a batch shares the same sensitivity when ordering, or tell us which value you need. This page will not guess that "the 3 t step must be 3.0" — the documentation does not say. The same two-value pattern also appears on HY-SQB, so it is not unique to this model.

What is it, and how does it relate to HY-SBS-M?

Short answer: the capacity range is the same (0.5/1/2/3 t) and both documents claim a "self-stabilising structure". But the wiring and the protection rating differ — this is not the same data sheet renamed.

HY-HLJ-M vs HY-SBS-M (both pages of documentation)
Item HY-HLJ-M (this page) HY-SBS-M
Capacity 0.5 to 3 t, 4 steps 0.5 to 3 t, 4 steps (same)
Wiring 4-wire (no sense) 6-wire with sense
Protection IP68 IP67
Sensitivity two values (2 / 3.0) single 2±0.002
Cable 5 m 3 m
Shared both self-stabilising, alloy steel nickel-plated, -20 to +60 °C, 381/350 Ω same

⚠️ Two points: ① 4-wire has no sense pair, and at 5 m the excitation drop matters more than on a short cable; ② IP68 is continuous immersion protection, steadier than IP67 on washdown — that is this model's main advantage over HY-SBS-M.

How is it wired?

Short answer: 4-wire. The text and the label on the unit agree.

4-wire colour code (text and unit label agree)
Colour Terminal Function
RedEXC+ / +input+ excitation
BlackEXC− / −input− excitation
GreenSIG+ / +output+ signal
WhiteSIG− / −output− signal
thick yellowSHIELDshield

⚠️ This is 4 wires plus one shield: the shield is not counted in "4-wire", it is for interference. Earth it at one end only. At 5 m this is a relatively long 4-wire run, so assess the drop.

What are the specifications?

Short answer: the company documentation gives 19 parameters, transcribed below. The sensitivity row carries two values — see the section above.

HY-HLJ-M specifications (company documentation)
Parameter Unit Value
SensitivitymV/V2±0.002 / 3.0±0.002
Non-repeatability%R.C.≤0.01
Hysteresis%R.C.≤0.02
Non-linearity%R.C.≤0.02
Creep (30 min)%R.C.≤0.02
Zero output%R.C.≤1
Compensated temperature range°C-10 to +40
Operating temperature range°C-20 to +60
Temperature effect on sensitivity%Load/°C≤0.002
Temperature effect on zero%R.C./°C≤0.002
Input resistanceΩ381±4
Output resistanceΩ350±1
Insulation resistanceMΩ≥5000 (50 V DC)
Recommended excitationV (DC/AC)5 to 15
Maximum excitationV (DC/AC)20
Safe overload%R.C.150
Ultimate overload%R.C.300
Protection—IP68
Cable lengthm5

What are the dimensions? ⚠️ The drawing's capacity groups differ from the product page

Short answer: the drawing carries a readable capacity table, but its capacity groups do not match the product page: the page lists 0.5/1/2/3 t, while the table lists "0.25/0.5/1 t, 2 t, 3 t, 5 t" — including 0.25 t and 5 t, which the page does not. The table is reproduced as drawn, and the discrepancy is stated.

Capacity-grouped dimensions (as drawn; capacity column in kg, other columns unmarked)
Capacity / kg A B D E F G H K M N P Q S
0.25 / 0.5 / 1 t 133.416.225.416.430.73057.9174.812.72215.93.17.7
2 t 138.716.225.416.730.835.857.9239.512.72215.93.111.7
3 t 139.719.225.416.73935.657.9239.5162215.9——
5 t 171.595.238.116.742.94373.829.4111934.922.23.111.7

⚠️ Three things to be clear about: ① the drawing has 0.25 t and 5 t groups that the product page does not list — an inconsistency, transcribed as-is with nothing added or removed; ask us if you need those steps; ② the Q and S cells of the 3 t row are blank on the drawing, not omissions here; ③ some digits are heavy in the original (e.g. column B on the 3 t row), so values are transcribed as legible — verify against the original drawing before machining. Column A (overall length) runs 133.4 to 171.5, a wide spread; use the row for your capacity.

What does it look like?

Short answer: five images are supplied — product photographs with a wiring label, and one dimension drawing with the capacity table. Two are used here.

HY-HLJ-M load cell: natural alloy-steel bar body with mounting holes, a wiring label on the side (EXC+ green, EXC- black, SIG+ white, SIG- red, SHIELD thick yellow); factory watermark
The label spells out the four wires plus shield. Company documentation.
HY-HLJ-M dimension drawing: side and top views with callouts, plus a capacity-grouped 14-column table (0.25/0.5/1 t, 2 t, 3 t, 5 t)
Dimension drawing (capacity table reproduced above; its groups differ from the product page). Company documentation.

⚠️ The label writes the shield as "SHIELD (Thick Yellow)". Across this factory's documents the shield has been written as "thick yellow", "long yellow" and pale-yellow — all the same thing; the actual wire governs.

Where does it fit, and where does it not?

Short answer: the documented applications are ground-level platforms and batching in chemical, food and pharmaceutical plants, 0.5 to 3 t.

The equipment and industries named above are the documented applications — where this cell goes — not a claim that this factory builds complete platforms.

How can I tell whether the cell has failed?

Short answer: input across red/black 381±4 Ω, output across green/white 350±1 Ω, insulation to body ≥5000 MΩ.

Three checks that can be done on site
Step What to do What to look for
1 Power off; measure red / black Around 381±4 Ω; open circuit means a broken internal wire
2 Measure green / white Around 350±1 Ω
3 Measure insulation to body MΩ range (≥5000); a clear drop means moisture or a damaged seal

Resistance readings cannot tell you whether the sensitivity is 2 or 3.0 — the resistances are identical. Check the nameplate or shipping documentation.

Where do the facts come from?

Both inconsistencies — the dual sensitivity value and the drawing's extra capacity groups — are transcribed unchanged.

Still unsure which step fits?

Tell us: load per cell and cell count, indicator model (4-wire or 6-wire), whether a specific sensitivity is required, washdown conditions, and mounting space and hole pattern.