⚠️ First, something the documentation leaves open: why two sensitivity values?
Short answer: the company documentation puts two values in the sensitivity cell — 2±0.002 and 3.0±0.002 — and does not say which capacity step gets which. Both are transcribed; this page does not assign them.
| Parameter | Unit | As written |
|---|---|---|
| Sensitivity | mV/V | 2±0.002 / 3.0±0.002 |
⚠️ This has a practical consequence: cells with different sensitivity share the load proportionally when paralleled, not equally. Mixing the two values on one scale makes calibration harder. Confirm that a batch has the same sensitivity when ordering, or tell us which value you need. This page will not guess that "the 5 t step must be 3.0" — the documentation does not say so.
What is it, and how does it relate to HY-SBS?
Short answer: features and most parameters match (low profile, alloy steel nickel-plated, IP67, 4-wire, 381/350 Ω, 3 m cable, -20 to +60 °C). The differences are the number of capacity steps and the sensitivity.
| Item | HY-SQB (this page) | HY-SBS |
|---|---|---|
| Capacity steps | 5 steps (0.5/1/2/3/5 t) | 4 steps (0.5/1/2/3 t, no 5 t) |
| Sensitivity | 2±0.002 / 3.0±0.002 (two values) | 2±0.002 (single value) |
| Features | 6 lines (same text as SBS) | 6 lines (same) |
| Wiring / resistance / cable | 4-wire / 381±4, 350±1 / 3 m | Identical |
| Outline | Drawing has a capacity table (0.5–2 t / 3–5 t groups) | Drawing has an L8 table (18/16) |
⚠️ This model also has a 100-300 kg version under the same name — different page, different parameter table. Buying the wrong range is the most common mistake with same-name multi-range products; confirm capacity before ordering.
How is it wired?
Short answer: 4-wire. The text and the drawing agree on the four conductors; the shield is where they diverge.
| Colour | Terminal | Function |
|---|---|---|
| Red | +input | + excitation |
| Black | −input | − excitation |
| Green | +output | + signal |
| White | −output | − signal |
| Shield | SHIELD | three statements differ: text omits it / drawing says pale-yellow / label on the unit says (Thick Black) |
⚠️ On the shield, three sources disagree and all three are transcribed: ① the product page text lists only four wires and does not mention a shield; ② the drawing marks pale-yellow; ③ the label on the unit says SHIELD (Thick Black). A sibling model (HY-SBT) writes Thick Yellow on its label — so the shield colour differs between batches and models. Work by the wire in hand and the shipping wiring diagram; never carry a colour over from another model. Earth the shield at one end only.
What are the specifications?
Short answer: the company documentation gives 19 parameters, transcribed below. The sensitivity row carries two values — see the section above.
| Parameter | Unit | Value |
|---|---|---|
| Sensitivity | mV/V | 2±0.002 / 3.0±0.002 |
| Non-repeatability | %R.C. | ≤0.01 |
| Hysteresis | %R.C. | ≤0.02 |
| Non-linearity | %R.C. | ≤0.02 |
| Creep (30 min) | %R.C. | ≤0.02 |
| Zero output | %R.C. | ≤1 |
| Compensated temperature range | °C | -10 to +40 |
| Operating temperature range | °C | -20 to +60 |
| Temperature effect on sensitivity | %Load/°C | ≤0.002 |
| Temperature effect on zero | %R.C./°C | ≤0.002 |
| Input resistance | Ω | 381±4 |
| Output resistance | Ω | 350±1 |
| Insulation resistance | MΩ | ≥5000 (50 V DC) |
| Recommended excitation | V (DC/AC) | 5 to 15 |
| Maximum excitation | V (DC/AC) | 20 |
| Safe overload | %R.C. | 150 |
| Ultimate overload | %R.C. | 300 |
| Protection | — | IP67 |
| Cable length | m | 3 |
What are the dimensions?
Short answer: the drawing carries a capacity-grouped dimension table (groups 0.5–2 t and 3–5 t), but the type in that table is too small to read reliably even enlarged. This page states that the table exists and does not copy the numbers.
| Group | Columns visible |
|---|---|
| 0.5–2 t | Rated capacity, L, L1, L2, L3, L4, H1, H2, H3, D1, D2 |
| 3–5 t | Same 11 columns |
⚠️ Why the numbers are not copied: on the original drawing the type is only a few pixels high; enlarged, the strokes merge and digits cannot be told apart (for example "82.7" versus "82.1"). Copying a misread dimension is worse than copying none — a base plate drilled to a wrong spacing is scrap. Use the original drawing or the shipping documentation for the values.
What does it look like?
Short answer: six images are supplied — product photographs, one dimension drawing and one wiring diagram. This page uses one photograph plus both drawings.
⚠️ Archive note: these images sit in the "product photo" folder on the old site, but two are actually drawings. This page uses them by what they are.
Where does it fit, and where does it not?
Short answer: the documented applications are ground-level platforms and building-material batching control, 0.5 to 5 t.
- Fits: ground-level platforms and building-material lines; 0.5 to 5 t per cell (one step more than HY-SBS, which stops at 3 t).
- Capacity limit: 0.5 to 5 t. The same model name also covers 100-300 kg — do not order the wrong range.
- ⚠️ Confirm sensitivity: two values are documented (2±0.002 / 3.0±0.002). Cells paralleled on one scale must share the same sensitivity, or the output splits proportionally and calibration gets harder.
- Temperature limit: -20 to +60 °C.
- Not covered: the 100–300 kg range (other version of this model); small bench platforms (single point); vessel batching (column modules); high-capacity platforms (double shear beam).
How can I tell whether the cell has failed?
Short answer: input across red/black 381±4 Ω, output across green/white 350±1 Ω, insulation to body ≥5000 MΩ.
| Step | What to do | What to look for |
|---|---|---|
| 1 | Power off; measure red / black | Around 381±4 Ω; open circuit means a broken internal wire |
| 2 | Measure green / white | Around 350±1 Ω |
| 3 | Measure insulation to body | MΩ range (≥5000); a clear drop means moisture or a damaged seal |
Resistance readings cannot tell you whether the sensitivity is 2 or 3.0 — the resistances are the same. Check the nameplate or the shipping documentation for sensitivity.
Where do the facts come from?
- Company product documentation: six features, 4-wire colour code, 5 capacity steps, 19 technical parameters (including the two sensitivity values).
- Company images (six): product photographs, a dimension drawing with the capacity-grouped table (values not copied), and a wiring diagram.
- Check against HY-SBS: both pages compared (capacity steps and sensitivity differ; the rest matches).
- Public technical literature: how paralleled cells split load by sensitivity, mounting requirements for batching.
The sensitivity is stated as two values with no mapping to capacity steps; both are transcribed and none is assigned.
Still unsure which step fits?
Tell us: hole pattern and thread of the existing base, load per cell and cell count, mounting height, indicator model, whether a specific sensitivity is required, and the temperature range.