Why is its resistance the highest here, and why check the indicator?
Short answer: input 1160±10 Ω, output 1000±3 Ω — two to three times the other models on this site (most are 350–772 Ω). That is not a typo; it is a common choice for large column cells, but it places real requirements on the indicator.
| Model | Input | Output |
|---|---|---|
| HY-GD-M (this page) | 1160±10 Ω | 1000±3 Ω |
| HY-SBD | 772±4 Ω | 700±1 Ω |
| HY-QS | 750±10 Ω | 703±2 Ω |
| HY-1022 / HY-1241 / HY-SSP1022 / HY-SSP1241 and similar | 400±10 Ω | 350±4 Ω |
| HY-SB-M / HY-SBC-M / HY-SBH-M and similar | 381±4 Ω | 350±1 Ω |
⚠️ Why this matters: a higher cell output resistance means more voltage division at the indicator input and more noise pickup. When selecting or replacing, confirm: ① the indicator's input impedance is far above 1000 Ω (most weighing indicators are 10 MΩ and up, but older or special units should be checked); ② do not parallel it with low-resistance models (381/350) — the output splits by resistance and calibration becomes difficult; ③ the fault-check reference here is 1160 / 1000 — borrowing 381/350 will fail a good cell.
What is it, and how does it split from the other high-capacity models?
Short answer: this is a column; HY-SBD and HY-QS are double shear beams. The three differ in outline and mounting: a column is a cylinder loaded through top and bottom bosses; a double shear beam is a long bar fixed at both ends and loaded in the middle.
| Model | Type | Capacity | Resistance (in/out) | Protection |
|---|---|---|---|---|
| HY-GD-M (this page) | column | 15–100 t (5 steps) | 1160 / 1000 | IP68 |
| HY-SBD | double shear beam | 15–50 t (5 steps) | 772 / 700 | IP67 |
| HY-QS | double shear beam | 15–50 t (5 steps) | 750 / 703 | IP67 (5 t) / IP68 (10-50 t) |
⚠️ Three rules: ① above 50 t (up to 100 t) only this model; ② IP68 at capacities up to 100 t — this is the only choice; ③ column vs double shear beam follows the mounting: a column suits vertical loading through aligned top and bottom bosses (as in a vehicle-scale load bearing), a double shear beam suits a long platform fixed at both ends. The documentation also claims "low profile, quick to install", an advantage in retrofits.
How is it wired?
Short answer: 6-wire. The text and the drawing agree wire for wire.
| Colour | Function | Terminal name |
|---|---|---|
| Green | + excitation | Excitation(+) |
| Black | − excitation | Excitation(−) |
| White | + signal | Signal(+) |
| Red | − signal | Signal(−) |
| Yellow | + sense | Sense(+) |
| Blue | − sense | Sense(−) |
| pale-yellow | shield | Shield |
⚠️ Here it is green = excitation+, white = signal+, red = signal−, not the 4-wire HY-1241 order (red = excitation+), so never carry a colour order across models. The cable is 13 m — a long run; this model has a sense pair, and if the sense wires are not connected the 13 m drop goes straight into the reading. Earth the shield at one end only.
What are the specifications?
Short answer: the company documentation gives 19 parameters, transcribed below. Note non-linearity ≤0.01, tighter than the ≤0.02 of most siblings.
| Parameter | Unit | Value |
|---|---|---|
| Sensitivity | mV/V | 2±0.002 |
| Non-repeatability | %R.C. | ≤0.01 |
| Hysteresis | %R.C. | ≤0.02 |
| Non-linearity | %R.C. | ≤0.01 |
| Creep (30 min) | %R.C. | ≤0.02 |
| Zero output | %R.C. | ≤1 |
| Compensated temperature range | °C | -10 to +40 |
| Operating temperature range | °C | -40 to +65 |
| Temperature effect on sensitivity | %Load/°C | ≤0.002 |
| Temperature effect on zero | %R.C./°C | ≤0.002 |
| Input resistance | Ω | 1160±10 |
| Output resistance | Ω | 1000±3 |
| Insulation resistance | MΩ | 5000 (50 V DC) |
| Recommended excitation | V (DC/AC) | 5 to 15 |
| Maximum excitation | V (DC/AC) | 20 |
| Safe overload | %R.C. | 150 |
| Ultimate overload | %R.C. | 300 |
| Protection | — | IP68 |
| Cable length | m | 13 |
The insulation cell reads "5000" with no "≥" sign in the original; it is transcribed as written.
What are the dimensions?
Short answer: the drawing carries symbols only, no numbers. It labels parts A, B, C, D, E, F and ΦA, ΦC, Φ40, Φ50 — without giving the value behind any symbol; the numbers have to be looked up per capacity.
| Symbol | Part |
|---|---|
| E | overall height (between top and bottom bosses) |
| F | top / bottom boss height |
| ΦA | body outside diameter |
| B / D / ΦC | bottom width / bottom nut across flats / top diameter |
| Φ40 / Φ50 | side boss diameter / centre bore |
⚠️ Why no numbers: the drawing has symbols but no value table. This page will not estimate — dimensions are used for machining and alignment, and an estimated number moves the risk onto the customer. Ask us for the drawing of this model and state the exact capacity (15/20/30/50/100 t).
What does it look like?
Short answer: three images are supplied — one product photograph, one wiring diagram and one dimension drawing (symbols only). All three are used here.
Where does it fit, and where does it not?
Short answer: the documented applications are vehicle weighing, rail weighing and heavy batching, 15 to 100 t.
The equipment types named above are the documented applications — where this cell goes — not a claim that this factory builds complete platforms.
- Fits: large vehicle-weighing platforms and rail weighing; 100 t class heavy duty; outdoor or wet sites needing IP68; retrofits with limited height.
- Why a column: a cylinder loaded through aligned top and bottom bosses is single-direction and easy to align, suited to load-bearing seat designs; the documentation also claims low profile and quick installation.
- ⚠️ Three checks: ① indicator input impedance (the cell output is 1000 Ω); ② do not parallel with low-resistance models; ③ dimensions must be requested (symbols only on the drawing).
- Not covered: up to 50 t where space allows (HY-SBD / HY-QS — lower resistance, easier to match); double-shear-beam mounting (HY-SBD); small capacities (single point or column modules).
How can I tell whether the cell has failed?
Short answer: input across green/black 1160±10 Ω, output across white/red 1000±3 Ω, insulation to body 5000 MΩ (as written, no "≥").
| Step | What to do | What to look for |
|---|---|---|
| 1 | Power off; measure green / black | Around 1160±10 Ω; open circuit means a broken internal wire |
| 2 | Measure white / red | Around 1000±3 Ω |
| 3 | Measure insulation to body | 5000 MΩ range; a clear drop means moisture or a damaged seal |
⚠️ Do not borrow HY-SBD (772/700) or HY-QS (750/703) references — this model is 1160/1000, and the wrong reference will fail a good cell. On a parallel bank, disconnect the other cells before measuring one.
Where do the facts come from?
- Company product documentation: six features, 6-wire colour code, 5 capacity steps, 19 technical parameters.
- Company images (three): product photograph, wiring diagram, dimension drawing (symbols only).
- Check against the other high-capacity models: HY-SBD and HY-QS pages (type, resistance and protection differ).
- Public technical literature: column versus double shear beam load paths, and what a high cell output resistance asks of the indicator input.
Dimensions are omitted because the drawing shows symbols without a value table.
Still unsure which step fits?
Tell us: load per cell and cell count, mounting form (load-bearing seat or long platform), indicator model (its input must suit a 1000 Ω output), whether the site stands wet, and any height limit.