Why is -40 °C the selection line for this model?
Short answer: the operating range is -40 to +65 °C. Most siblings start at -20 °C (one at -30 °C); this one pushes the floor to -40 °C, and that is what separates it in selection.
| Model | Operating range |
|---|---|
| HY-SBC-M (this page) | -40 to +65 °C |
| HY-SBD | -40 to +65 °C |
| HY-QS | -30 to +70 °C |
| HY-SBS / HY-SBT / HY-SQB | -20 to +60 °C |
| HY-HLJ-M / HY-0743-M / HY-0745A-M | -20 to +60 °C |
⚠️ Two directions to check: ① cold sites (cold stores, outdoor winter, unheated workshops) — here -40 °C is a hard requirement and a -20 °C model will not do; ② but the upper limit is only +65 °C, below HY-QS at +70 °C, so hot processes need separate checking. Look at both ends, not just one.
What is it, and how does it differ from HY-SBS-M?
Short answer: the same capacity range (0.5/1/2/3 t), both documented as "self-stabilising", both IP67. What differs is the wiring (4-wire vs 6-wire) and the operating range.
| Item | HY-SBC-M (this page) | HY-SBS-M |
|---|---|---|
| Capacity | 0.5 to 3 t, 4 steps | 0.5 to 3 t, 4 steps (same) |
| Wiring | 4-wire (no sense) | 6-wire with sense |
| Operating range | -40 to +65 °C | -20 to +60 °C |
| Protection | IP67 | IP67 (same) |
| Cable | 3.1 m | 3 m |
| Shared | self-stabilising, alloy steel nickel-plated, 381/350 Ω, 150%/300% | same |
Selection: cold duty takes this model (-40 °C); sense-wire compensation takes HY-SBS-M (6-wire). They cannot both be had at once — this model has no sense pair, HY-SBS-M only reaches -20 °C.
How is it wired?
Short answer: 4-wire. The text and the label on the unit agree.
| Colour | Terminal | Function |
|---|---|---|
| Red | EXC+ / +input | + excitation |
| Black | EXC− / −input | − excitation |
| Green | SIG+ / +output | + signal |
| White | SIG− / −output | − signal |
| thick yellow | SHIELD | shield |
⚠️ This is 4 wires plus one shield; the shield is for interference and is not counted in "4-wire". Earth it at one end only.
What are the specifications?
Short answer: the company documentation gives 19 parameters, transcribed below.
| Parameter | Unit | Value |
|---|---|---|
| Sensitivity | mV/V | 2±0.002 |
| Non-repeatability | %R.C. | ≤0.01 |
| Hysteresis | %R.C. | ≤0.02 |
| Non-linearity | %R.C. | ≤0.02 |
| Creep (30 min) | %R.C. | ≤0.02 |
| Zero output | %R.C. | ≤1 |
| Compensated temperature range | °C | -10 to +40 |
| Operating temperature range | °C | -40 to +65 |
| Temperature effect on sensitivity | %Load/°C | ≤0.002 |
| Temperature effect on zero | %R.C./°C | ≤0.002 |
| Input resistance | Ω | 381±4 |
| Output resistance | Ω | 350±1 |
| Insulation resistance | MΩ | ≥5000 (50 V DC) |
| Recommended excitation | V (DC/AC) | 5 to 15 |
| Maximum excitation | V (DC/AC) | 20 |
| Safe overload | %R.C. | 150 |
| Ultimate overload | %R.C. | 300 |
| Protection | — | IP67 |
| Cable length | m | 3.1 |
What are the dimensions?
Short answer: the drawing carries a 12-column capacity table. It is reproduced below, but some digits are heavy in the original — verify against the drawing before machining. Units are not marked.
| Capacity / t | M | L2 | L3 | L5 | G | D | E (overall) | M1 | D2 | K | N | P |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0.5 | 16 | 22 | 12.5 | 50.5 | 30.2 | 55 | 133.4 | 25.4 | 20.2 | 15.4 | 22 | 17 |
| 1 | 16 | 22 | 12.5 | 50.5 | 30.4 | 56.5 | 133 | 25.4 | 20.2 | 16.4 | 22.7 | 17 |
| 2 | 16 | 22 | 14.7 | 50.2 | 30.4 | 56.5 | 137 | 25.4 | 20.2 | 16.4 | 22 | 17 |
| 3 | 16 | 22 | 14.7 | 50.2 | 30.4 | 56.5 | 139 | 25.4 | 20.2 | 16.4 | 22 | 17 |
⚠️ Two points: ① the rows are not all the same — E (overall length) climbs from 133.4 to 139, L3 changes from 12.5 to 14.7 at 2 t and above, D from 55 to 56.5 — use the row for your capacity; ② some digits are heavy (e.g. column N on the 1 t row), transcribed as legible and to be verified against the original drawing before machining.
What does it look like?
Short answer: six images are supplied — product photographs, more than one wiring-label view, and one dimension drawing with the capacity table. Two are used here.
⚠️ The label writes the shield as "SHIELD (Thick Yellow)". Across this factory's documents it has appeared as "thick yellow", "long yellow" and pale-yellow — all the same thing; the actual wire governs.
Where does it fit, and where does it not?
Short answer: the documented applications are ground-level platforms and building-material batching, 0.5 to 3 t.
The equipment and industries named above are the documented applications — where this cell goes — not a claim that this factory builds complete platforms.
- Fits: medium-to-small batching; cold sites (cold stores, outdoor winter, unheated workshops) — this model's main advantage here.
- Why -40 °C: most siblings stop at -20 °C, and materials and seals behave differently in the cold; that is a hard selection criterion.
- ⚠️ Two limits: ① the upper limit is +65 °C only (HY-QS reaches +70 °C); ② 4-wire without sense, so sense compensation needs a 6-wire model.
- Not covered: duties needing a sense pair (HY-SBS-M); hot processes (HY-QS, +70 °C); IP68 (HY-HLJ-M / HY-SSH).
How can I tell whether the cell has failed?
Short answer: input across red/black 381±4 Ω, output across green/white 350±1 Ω, insulation to body ≥5000 MΩ.
| Step | What to do | What to look for |
|---|---|---|
| 1 | Power off; measure red / black | Around 381±4 Ω; open circuit means a broken internal wire |
| 2 | Measure green / white | Around 350±1 Ω |
| 3 | Measure insulation to body | MΩ range (≥5000); a clear drop means moisture or a damaged seal |
Where do the facts come from?
- Company product documentation: six features, 4-wire colour code, 4 capacity steps, 19 technical parameters.
- Company images (six): product photographs with a wiring label, and a dimension drawing with the 12-column capacity table.
- Operating-range check against models here: the other product pages.
- Public technical literature: how low temperatures affect cell materials and seals, 4-wire versus 6-wire over long runs.
Values are transcribed from company documentation and are not a commitment for a specific project.
Still unsure which step fits?
Tell us: load per cell and cell count, the site's lowest and highest temperatures, indicator model (4-wire or 6-wire), whether immersion protection is needed, and mounting space and hole pattern.