How does it differ from HY-1241?
Short answer: in four ways, none of them minor. ① wire count (6 vs 4); ② red and green mean the opposite; ③ overall length (150 vs 174); ④ how the protection rating is stated (this page says IP67 in both places; HY-1241 says IP65 in its feature line and IP67 in its table).
| Item | HY-SSP1241 (this page) | HY-1241 |
|---|---|---|
| Wiring | 6-wire with sense | 4-wire without sense |
| Red / green | red = +signal, green = +excitation | red = +excitation, green = +signal |
| Capacity | 30 to 250 kg, 5 steps | 30 to 250 kg, 5 steps (same) |
| 19 parameters | Item-for-item same | Item-for-item same |
| Protection | IP67 (consistent) | IP65 (features) / IP67 (table) — explained on that page |
| Overall length | 150 | 174 |
| Cable | 2 m | 2 m (same) |
⚠️ The second row matters most: red and green are swapped between the two. Wiring this cell by HY-1241 habit (red as excitation) puts excitation onto the signal pair and signal onto the excitation pair — at best no reading or a reversed one, at worst damage. Re-check the colour order from this page whenever you change model; never carry an old harness over.
How is it wired, and why are red and green reversed?
Short answer: 6-wire, and the text agrees with the drawing wire for wire. The colour assignment really is different from HY-1241 — that is what each model's documentation specifies, not a typo, and this page transcribes it as written. The wire in hand and the shipping diagram govern.
| Colour | Function | Terminal name |
|---|---|---|
| Green | + excitation | Excitation(+) |
| Black | − excitation | Excitation(−) |
| Red | + signal | Signal(+) |
| White | − signal | Signal(−) |
| Blue | + sense | Sense(+) |
| Yellow | − sense | Sense(−) |
| pale-yellow | shield | Shield |
| Colour | HY-SSP1241 | HY-1241 |
|---|---|---|
| Red | + signal | + excitation |
| Green | + excitation | + signal |
| Black | − excitation | − excitation |
| White | − signal | − signal |
⚠️ The right way to change model: ① wire by terminal name (Excitation / Signal / Sense), not by colour; ② cells sharing one indicator should share the same wire count and colour order; ③ earth the shield at one end only.
What are the specifications?
Short answer: the company documentation gives 19 parameters, identical to HY-1241, transcribed below.
| Parameter | Unit | Value |
|---|---|---|
| Sensitivity | mV/V | 2±0.2 |
| Non-repeatability | %R.C. | ≤0.01 |
| Hysteresis | %R.C. | ≤0.02 |
| Non-linearity | %R.C. | ≤0.02 |
| Creep (30 min) | %R.C. | ≤0.02 |
| Zero output | %R.C. | ≤10 |
| Compensated temperature range | °C | -10 to +40 |
| Operating temperature range | °C | -20 to +65 |
| Temperature effect on sensitivity | %Load/°C | ≤0.002 |
| Temperature effect on zero | %R.C./°C | ≤0.002 |
| Input resistance | Ω | 400±10 |
| Output resistance | Ω | 350±4 |
| Insulation resistance | MΩ | ≥5000 (50 V DC) |
| Recommended excitation | V (DC/AC) | 5 to 15 |
| Maximum excitation | V (DC/AC) | 20 |
| Safe overload | %R.C. | 150 |
| Ultimate overload | %R.C. | 300 |
| Protection | — | IP67 |
| Cable length | m | 2 |
What are the dimensions?
Short answer: one set of callouts with overall length 150 (HY-1241 is 174, so the same base plate will not fit). Units are not marked and none are added here.
| View | Values shown |
|---|---|
| Side | Overall 150, height 38 |
| Top | Width 20.4, 5.2, 25.4 (repeated), 39.8; holes 4-Φ8.5 |
⚠️ Against the HY-1241 drawing: that one is overall 174, height 65, width 60, hole spacing 122/19/19/7; this one is overall 150, height 38, width 20.4. The numbers are far apart — the two cannot share a base plate hole pattern.
What does it look like?
Short answer: four images are supplied — one product photograph, one dimension drawing and one wiring diagram. This page uses one photograph plus both drawings.
⚠️ Archive note: these images sit in the "product photo" folder on the old site, but two are actually drawings. This page uses them by what they are.
Where does it fit, and where does it not?
Short answer: the documented applications are bench scales, packing scales and process weighing, 30 to 250 kg. The duty is the same as HY-1241; what differs is the wiring and the outline.
- Fits: medium bench platforms, packing scales, process weighing; indicators wired for 6-wire, or where sense wires are wanted to absorb excitation drop.
- Why 6-wire: on long runs or with a 6-wire indicator, the sense pair compensates the drop. The 4-wire HY-1241 cannot do that.
- ⚠️ Changing model: red and green are swapped and the overall length differs (150 vs 174). Wire by terminal name, not colour; re-check the base-plate holes.
- Capacity limits: 30 to 250 kg. Below that see HY-SSP1022 (3 to 30 kg); larger platforms see HY-SSP-IL (150 to 2000 kg).
- Not covered: indicators that only accept 4-wire; vessel batching (column modules); high-capacity platforms (HY-SBD).
How can I tell whether the cell has failed?
Short answer: input across green/black 400±10 Ω, output across red/white 350±4 Ω, insulation to body ≥5000 MΩ.
| Step | What to do | What to look for |
|---|---|---|
| 1 | Power off; measure green / black | Around 400±10 Ω; open circuit means a broken internal wire |
| 2 | Measure red / white | Around 350±4 Ω |
| 3 | Measure insulation to body | MΩ range (≥5000); a clear drop means moisture or a damaged seal |
⚠️ Note which pairs are measured here: green/black for excitation and red/white for signal — the reverse of HY-1241 (where red/black is excitation and green/white is signal). Probing the wrong pair gives numbers that mislead.
Where do the facts come from?
- Company product documentation: five features, 6-wire colour code, 5 capacity steps, 19 technical parameters.
- Company images (four): one product photograph, a dimension drawing, a wiring diagram.
- Item-by-item check against HY-1241: both pages and both drawings (wiring, colour order and outline differ; parameters and capacities do not).
- Public technical literature: 4-wire versus 6-wire over long runs, corner compensation in single-point cells.
Values are transcribed from company documentation and are not a commitment for a specific project.
Still unsure which step fits?
Tell us: whether the indicator is 4-wire or 6-wire, platform size and construction, load per cell and cell count, the existing base-plate hole pattern (this model differs from HY-1241), and whether there is washdown.